ARTICLE |
a Division of Postmarket Surveillance, Office of Surveillance and Biometrics
b Division of Anesthesiology, General Hospital, Infection Control and Dental Devices, Office of Device Evaluation, Center for Devices and Radiological Health, Food and Drug Administration, Rockville, Maryland
| ABSTRACT |
|---|
|
|
|---|
METHODS. A search for medical device adverse event reports from January 1, 1996 through December 31, 2005, involving insulin pumps or patient-controlled analgesic pumps used by patients who were aged 12 to 21 years was conducted in the Food and Drug Administration's Manufacturer and User Facility Device Experience Database. Reports were reviewed for demographic characteristics, type of adverse event, and patient morbidity, and potential contributory factors were classified from narratives in the reports.
RESULTS. A total of 1674 reports were identified: 1594 for insulin pumps and 53 for patient-controlled analgesic pumps. In reports of insulin pump events, there were 13 reported deaths, 2 reports that indicated possible suicide attempts, and several additional reports indicating severe hypoglycemic or hyperglycemic events that seemed to be device-related. A total of 102 (6.4%) insulin-pump reports highlighted factors that may have contributed to the adverse event, including problems associated with compliance, education, sports-related activities, and dropping or damaging the pump. Eighty-two percent of cases involving the insulin pump resulted in hospitalization. Half of the reports involving patient-controlled analgesic pumps indicated that the patient received an excess of medication; tampering and noncompliance were evident in some cases.
CONCLUSIONS. Adolescents are a special population who deserve careful consideration of risk and benefit for use of device technology. Studies need to further identify safety problems in this age group.
Key Words: medical devices adolescent infusion pump diabetes analgesia adverse events postmarket surveillance
Abbreviations: PCA—patient-controlled analgesia FDA—Food and Drug Administration PMA—premarket approval DKA—diabetic ketoacidosis MAUDE—Manufacturer and User Facility Device Experience
Infusion pumps are a growing technology that has been used predominately in the adult population but that has in recent years become more prevalent among adolescents. Large population-based studies have shown high rates of preventable adverse events among hospitalized children,1 notably adolescents.2 Miller et al1 studied pediatric adverse events by using patient safety indicators and found that both young adolescents, aged 10 to 14, and older adolescents, aged 15 to 19, experienced the highest rates of in-hospital adverse events, with older adolescents experiencing nearly twice the adverse event rate of younger group (odds ratio: 1.9; 95% confidence interval: 1.7–2.0). Adolescents who have diabetes may experience device-related problems for a variety of reasons that vary with patient and disease characteristics.3 Few studies have examined device-related problems for patient-controlled analgesia (PCA) pumps.
From January 1, 2005, through December 31, 2005, the Food and Drug Administration (FDA) received 5 death reports in which teenage insulin pump users may have been negligent or noncompliant with the use of the device. This prompted a review of infusion pump–related adverse events in this age group to understand better the scope, severity, and type of circumstances described in reports.
It is important to understand the regulatory background and context for which pump devices are used. The majority of insulin pumps that are used for treatment of diabetes are cleared by the FDA through the 510(k) or premarket notification process. The 510(k) process is one type of regulatory pathway for medical devices to enter the market. It allows a manufacturer to show substantial equivalence to a previously approved and marketed device.4 Alternatively, marketing approval may be obtained by demonstrating with reasonable assurance that the device is safe and effective for its intended use through premarket approval (PMA).5 Many pumps are cleared for use in the general population through the 510(k) process; however, some pumps, such as implantable pumps, are approved through the PMA process. The PMA is the most stringent type of device marketing application required by FDA.
An insulin pump may offer tighter control of blood glucose levels while aiming to avoid diabetic ketoacidosis (DKA) and hypoglycemia. Pumps can be programmed to provide insulin at a basal rate and also to deliver mealtime or supplementary bolus doses of insulin. The attainment of normoglycemic levels helps to reduce the risk of complications of diabetes; however, insulin pumps have the potential to overmedicate and undermedicate. For example, insulin delivery may be impeded by a weak or depleted battery, occlusion of the infusion set, or even leakage. In some cases, insulin may precipitate and occlude the needle or tubing, thereby impeding delivery of the drug. Other hazards may result in free-flow or excessive insulin administration, putting patients at risk for hypoglycemia.6
There are many device components, such as hardware and software, that can contribute to the delivery system and proper device function. If the device should malfunction, then this may result in overinfusion, underinfusion, or delivery of a lethal dose of insulin. Errors associated with filling, refilling, or changing dosage or concentration of the product delivered by the device may also lead to problems of overdelivery or underdelivery of drugs. Other critical issues that should be considered are the accuracy of the software (eg, telemetry), accuracy of dosage calculation, and accuracy of programming the device. Catheter-related problems, such as kinking, leakage, and blockage, may compromise patient outcome. In some implantable pumps, migration may occur, and blockage may also result from marked fibrosis and tissue growth at the implantation site, necessitating removal of the device.
The genesis for PCA pumps was a result of the quest for more effective pain treatment. PCA technology has evolved and is not restricted to a single class of analgesics or single mode of administration. The pumps deliver intravenous, subcutaneous, and epidural infusates. Patients do not have to wait for administration of their medications by a nurse and may self-administer narcotic analgesia as prescribed by their physician. The two most prevalent modes of administration found in PCA pumps are demand dosing, which delivers a fixed, self-administered dose intermittently, and a fixed-rate background continuous infusion with supplemental patient demand dosing. Most pumps offer both modes with programmable features that allow the pumps to be adjusted by clinicians, who can adjust the bolus size, lockout intervals, and continuous rate infusion. The greatest hazard to patients is increased risk for overmedication, particularly narcotic overinfusion that may lead to respiratory depression. PCA pumps differ in the methods used to store and deliver medication and hardware and software designs. Pumps can be worn, attached to an intravenous pole, or placed at the bedside.
| METHODS |
|---|
|
|
|---|
803.3). Information from these reports is entered into the FDA's MAUDE database. All reports were identified by using codes specific for insulin or PCA pumps, and duplicate reports were merged. Data were reviewed and evaluated for all reports by using the narrative text and the summary patient or device problem codes indicating the overall nature of the device or patient problem as described by the reporter. For example, with insulin pumps, a patient problem code indicating "hypoglycemia" or "low blood glucose" would be assigned to a report indicating that the patient received too much insulin and experienced a low blood sugar. When summarizing the total number of reports with hypoglycemic complications, we included reports that had more or less of the following patient problem codes: hypoglycemia, low blood glucose, overmedication, insulin shock, and overdose. Reports that were classified as hyperglycemia were identified by using codes for hyperglycemia, undermedication, and underdose. Severe complications as indicated by patient problem code or text indicating DKA, coma, and seizures were also assessed. Reports were further reviewed for the types of injury or malfunction. Details of patient problems in the narrative text allowed us to group further the types of adolescent issues into the following categories: education, noncompliance, sports and other activities, device misuse, risk-taking behaviors, responsibility and care of the device, pubertal growth, problems occurring while away from home, and lack of parental supervision. Reports were classified as "noncompliance" when the text described that the adolescent failed to keep parent(s) informed, failed to follow diabetic protocol and report problems to caregiver, failed to provide adequate feedback, and was negligent in caring for the device. Reports that were classified in the education category revealed that the adolescent was unfamiliar with the device and its operation and how to troubleshoot. Reports that were classified in the risk-taking category had narrative text describing instances of alcohol abuse, stress, and possible suicidal attempts.
Adverse event reports for PCA pumps were similarly reviewed. Complications that were associated with narcotic use and narcotic drug–related adverse effects were also evaluated.
| RESULTS |
|---|
|
|
|---|
|
|
A total of 167 (10.5%) reports included patient and device problems with hypoglycemia or overdelivery of insulin. Several reports were coded as overmedication (n = 15; 17.5%), overdose (n = 7; 4.2%), and insulin shock (n = 1; 0.6%). Device codes also indicated incorrect use of the device (n = 18), problem with self-activating key and overbolusing (n = 9), error message (n = 8), and alarm problems (n = 8). In 60 (35.9%) of the 167 reports, the device problem was coded as unknown.
In 102 (6.4%) reports, special adolescent issues were identified (Table 2). Eighty-two (80.4%) of the 102 reports resulted in hospitalization. The top 3 issues that were associated with these reports were education, noncompliance, and problems during sports or other activities. Four reports indicated risk-taking behaviors (including 2 reports of insulin overbolusing that were thought to be suicide attempts).
|
|
| DISCUSSION |
|---|
|
|
|---|
Although insulin pumps may mimic physiologic insulin replacement, they may be problematic because of adolescent lifestyle and psychosocial factors.3 Adolescent-aged issues were implicated in many of the FDA reports. We noted several adverse events related to extracurricular activities and accidental dropping and breaking of the pump. This was also reported by Burdick et al25 in a study of 48 adolescents who received insulin pump therapy: 52% reported disconnecting their pump for exercise. A Finnish study of adolescents with diabetes revealed that 75% believe that they had satisfactorily complied with the protocol, whereas, on review, only one fifth had actually complied.26 This study pointed out that teens tend to give health care providers answers that were compliant with expectations and identified conflicts between the adolescent and parents or other caregivers about diabetic care. Better diabetic control among adolescents has been associated with greater parental involvement.27 The FDA reports revealed 3 deaths that occurred at times when there was no parental supervision.
Difficulties may also arise when teens are careless or tamper with their medical devices, injuring themselves. An increased risk for hypoglycemic complications may occur with binge drinking of alcohol. In our study, a patient who was away at college was found dead in his dorm room after being out with friends the night before. Teenagers with chronic medical disorders have an increased risk for stress and behavioral and psychiatric disorders.28,29 Two hospitalized patients with diabetes in our series overbolused with insulin, and the reports indicated that these may have been suicide attempts. A review of the literature revealed a case report of a 14-year-old girl who administered hair conditioner through her intravenous infusion pump and died.30 It is unclear whether her death was attributable to risk-taking behavior or an intentional suicide. Healthcare providers should be aware of possible suicidal ideation among teens with diabetes and potential risks with misuse of medical devices.
Only a few studies have evaluated insulin pump malfunctions that occurred in adolescent patients. A retrospective study of 6 adolescents found 1 teen hospitalized on 2 different occasions for DKA associated with device malfunction.16 Weintrob et al31 conducted a small study of 20 adolescent patients and found a higher number of ketotic events for pump users (P = .003), all related to infusion set problems. In another study, Saha et al32 reported that 3 adolescent pump users experienced ketoacidosis related to disconnection and occlusion of the infusion catheters. Our study found that one third of all pump-related problems were malfunctions, and detailed information reported various device problems with error messages and problems with the alarm, catheter, and screen display.
PCA Pumps
PCA pumps have become an important strategy in pain management for pediatric medical conditions. Although the delivery of opioids has been shown to be effective for pain relief,33 their administration carries the risk for respiratory depression. In our study, we found 12 cases of respiratory depression that required medical intervention. Two studies in the literature also reported older children and adolescents who experienced problems with oxygen desaturation.34,35 In our series, more than half of the adverse events indicated that the adolescent received an excess bolus of opioid. In most of the reports, it was unclear whether incorrect dosage boluses, programming errors, drug errors, or accidental bolus administration had occurred. Although studies do indicate that age affects opioid dosing36 few have discussed the human factor issues surrounding the use in the adolescent age group. In a pediatric case series of PCA use among patients with sickle cell disease, Shapiro et al37 noted that 1 adolescent tampered with the device as a possible suicidal gesture. We found 2 cases in our series. Three deaths reported to the FDA indicated that the adolescents were found unconscious too late after overdelivery of narcotics, and other reports seemed to be near-miss occurrences that would have been lethal if they had gone unnoticed with time delay. Patients with PCA pumps may allow others to push the button on their pump, and this can result in serious consequences and death.38 This has led to advisories from manufacturers, ECRI Institute, and the Joint Commission on Accreditation of Healthcare Organizations that allowing anyone other than the patient to press the PCA delivery request button is contraindicated.39 Life-threatening complications of respiratory depression, oversedation and confusion highlight the need for careful selection and monitoring of adolescent patients who receive this device.
Limitations
There are many limitations to this study. The majority of adverse event reports were submitted by manufacturers. The accuracy and completeness of the reports were not verified by the FDA. By limiting our search to the specific age group (12–21 years), we were unable to capture reports with missing age information. Medical device adverse event reports cannot clearly establish a cause-and-effect relationship of an injury or illness with the device. The recognition of a device-related event is often subjective and imprecise. The reports had insufficient details about patient factors, such as patient compliance, socioeconomic status, medical history, disease severity, duration of disease, history of substance abuse, years of pump use, and developmental stage of puberty. There were no data on use of pump devices and sales in this age group to provide the denominator to calculate incidence rates of adverse events. Adverse event reports for PCA pumps also had missing information regarding the type of analgesia, dosage, diagnosis, and severity of illness.
| CONCLUSIONS |
|---|
|
|
|---|
| FOOTNOTES |
|---|
Address correspondence to Judith U. Cope, MD, MPH, Office of Pediatric Therapeutics, FDA, Parklawn Room 13B45 HFG-2, 5600 Fishers La, Rockville, MD 20850. E-mail: Judith.cope{at}fda.hhs.gov
The authors have indicated they have no financial relationships relevant to this article to disclose.
| What's Known on This Subject Infusion pumps are a growing technology that is becoming more widely used in the adolescent age group. Insulin pumps may offer a tighter control of blood glucose. Patient-controlled analgesia pumps deliver pain medications with continuous infusion and programmable features.
|
| What This Study Adds This study examines 10 years of FDA adverse event reports with the use of insulin and PCA pumps in the adolescent age group. Concerns about the use of this technology in the 12- to 21-year age group are discussed.
|
| REFERENCES |
|---|
|
|
|---|
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||