a Division of Neurology, Children's Hospital of Philadelphia, Philadelphia, Pennsylvania
b Departments of Neurology and Pediatrics, School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania
c Department of Biostatistics, University of Pittsburgh, Pittsburgh, Pennsylvania
| ABSTRACT |
|---|
|
|
|---|
OBJECTIVES. This study's purpose was to determine the sensitivity of neonatal seizure detection in a single electroencephalogram channel (C3
C4), used to simulate the raw signal from which aEEG is derived. We also aimed to determine the sensitivity of seizure detection by neonatologists by using aEEG and to establish those neonatal seizure characteristics that are associated with their correct detection by aEEG.
METHODS. Conventional electroencephalograms with neonatal seizures were reviewed for electroencephalogram background and neonatal seizure characteristics (site of onset, duration, and peak-to-peak amplitude). The presence, duration, and peak-to-peak amplitude of each seizure were simultaneously noted in a single electroencephalogram channel (C3
C4). aEEGs generated from this channel were reviewed for background and seizures by 6 neonatologists with varying aEEG interpretation expertise.
RESULTS. A total of 851 neonatal seizures from 125 conventional electroencephalograms were analyzed. The patients' conceptional ages were 34 to 50 weeks. Because 94% of the conventional electroencephalograms had
1 neonatal seizure visible in C3
C4, and 78% of all neonatal seizures appeared in the C3
C4 channel, the theoretical sensitivity of seizure detection in a single electroencephalogram channel was high. However, seizures were briefer and lower in amplitude in C3
C4 compared with conventional electroencephalography. Neonatologists identified seizures in 22% to 57% of the 125 records of neonatal seizure. They detected 12% to 38% of the 851 individual seizures. Multivariate analysis revealed that the appearance of seizures in C3
C4, neonatal seizure duration, seizure amplitude, seizure count per hour, and neonatologists' experience with aEEG interpretation all correlated with neonatal seizure detection.
CONCLUSIONS. Even among physicians who have extensive experience, many neonatal seizures are difficult to detect on an aEEG, especially when they are infrequent, brief, or of low amplitude.
Key Words: neonatal seizures neonatal EEG amplitude-integrated EEG cerebral function monitor
Abbreviations: EEG—electroencephalogram aEEG—amplitude-integrated electroencephalogram
Clinical neonatal seizures are diagnosed in 1.5 to 3.5 per 1000 live term births.1–3 Neonates who have seizures are at high risk for death or significant neurologic disability.4,5 On the basis of animal and limited human data, it is also believed that the corresponding electrographic seizures have an inherent adverse effect on neonates' neurodevelopmental outcome.6–8
Because neonates with seizures are at such high risk for mortality and neurodevelopmental abnormalities, neonatologists and neurologists have great interest in their early and accurate detection. In clinical practice, neonates at high risk for seizures are visually monitored for clinical signs of seizures. When there is a clinical suspicion of seizures, a routine electroencephalogram (EEG) is obtained, and empiric treatment is often administered. This approach presumes that most electrographic seizures give rise to visually observable clinical seizures. However, the majority of electrographic neonatal seizures are subclinical.5,9–11 Therefore, continuous EEG monitoring is required for prompt and reliable electrographic seizure detection.
In an attempt to identify electrographic seizures early in an at-risk neonate's course, cerebral function monitors, such as amplitude-integrated EEGs (aEEGs), have been introduced in many NICUs. The device uses a single EEG channel with biparietal electrodes (P3
P4) to monitor the EEG background and to detect seizures.12 The raw signal is highly filtered, processed, and compressed to display the envelope of lower and upper amplitudes of the EEG signal over time. The electrical background recorded by aEEG is reported to correlate well with conventional neonatal EEG.13–15 However, because few studies have adequately addressed this question,15,16 it is not yet clear how sensitive or specific aEEG is for electrographic seizure detection.
The purpose of this study was to characterize a large number of contemporary electrographic neonatal seizures by conventional EEG and to determine the sensitivity of neonatal seizure detection in a single EEG channel (C3
C4), used to simulate the raw signal from which aEEG is derived. We then examined the sensitivity of electrographic seizure detection by neonatologists using aEEG and sought to identify those seizure characteristics that were significantly associated with their detection by aEEG.
| METHODS |
|---|
|
|
|---|
Each conventional EEG was interpreted by 2 pediatric electroencephalographers (Drs Shellhaas and Clancy). Individual records were classified as having a normal or mildly, moderately, or markedly abnormal background, according to a standard classification system.17 An electrographic seizure was defined as a sudden, repetitive, evolving, and stereotyped ictal pattern with a clear beginning, middle, and ending, an amplitude of
2 µV, and a minimum duration of 10 seconds. To be counted as separate events, individual seizures had to be separated by a minimum of 10 seconds. Seizure onset and termination times, duration, electrode of onset, and maximal peak-to-peak amplitude were recorded.
A new EEG channel, C3
C4, was created to represent the raw data from which an aEEG would be derived. Historically, aEEG leads are placed by neonatologists over the biparietal region (P3
P4), the apex of the vascular watershed area. Because the locations of P3
P4 are not included in the international 10–20 system of electrode placement, modified for neonates, we chose their closest neighbors (C3
C4), to create a single-channel raw EEG from which the aEEG traces would be derived (Fig 1). In a typical term infant, C3 is only
4 cm anterior to P3. Thus, there is little chance of error resulting from substituting C3
C4 for P3
P4.
|
C4 channel. For each seizure, the electrode where the seizure originated and the times of seizure onset and termination in both the conventional EEG and the C3
C4 channel were recorded, as was the maximal ictal peak-to-peak amplitude. The interictal peak-to-peak amplitude was also measured for both the conventional EEG and the C3
C4 channel, and the ratio of ictal to interictal amplitudes was calculated for each seizure.
Finally, the single-channel raw EEG recordings were converted to aEEG traces. The durations of the conventional EEG, single channel EEG, and aEEG recordings were identical, because the latter 2 were derived directly from the original EEG traces. The aEEGs were interpreted by 6 neonatologists with varying aEEG interpretation expertise. All had
1 year of experience with aEEG, and 2 were internationally recognized experts. They each received written instructions and a packet of 144 paper aEEG traces. They were informed that some aEEGs contained seizures but others did not. The collection contained 125 records with seizures, which were randomly interspersed with 19 aEEGs without seizures. The neonatologists were asked to (1) score the technical quality of the tracings, (2) classify the aEEG background, and (3) mark with a pen the onset (O) and termination (T) of all portions of the aEEG traces that they believed represented electrographic seizures (Fig 2). The raw EEGs were not made available to the neonatologists. In clinical practice, most modern aEEG equipment can call up and display the single-channel raw EEG if requested by the aEEG interpreter. Therefore, the participating neonatologists were instructed to mark the areas of the aEEG traces that would raise their suspicion for a seizure and, consequently, would have triggered them to look at those sections of raw EEG traces had they been available.
|
2 statistics were used to compare categorical variables. Multivariate analysis was performed using generalized estimating equation with binary outcome (Stata 8; Stata Corp, College Station, TX).18,19 Variables examined in the model included the seizure's visibility in C3
C4, seizure duration (seconds), number of seizures per hour, location of seizure onset, conventional EEG background classification, ictal peak-to-peak amplitude (microvolts), and the neonatologist's level of expertise in aEEG interpretation. A P value of <.05 was considered to be statistically significant. | RESULTS |
|---|
|
|
|---|
C4, and the derived aEEG recordings were all equal and ranged from 23 to 145 minutes.
Eighty-one percent of all seizures (691 of 851) originated in the central (C3 or C4), temporal (T3 or T4), or midline vertex (CZ) electrodes (Table 1). Ninety-four percent (118 of 125) of the conventional EEGs had
1 seizure appearing in the single C3
C4 channel. Seventy-eight percent (664 of 851) of the individual seizures were visible in this channel. However, compared with conventional EEG, the seizures in C3
C4 were briefer (mean: 132 vs 100 seconds; P < .001), less frequent (mean: 7.0 vs 5.2 seizures per hour; P < .001), and lower in peak-to-peak amplitude (mean: 145 vs 111 µV; P < .001; Table 2). The ratio of ictal to interictal peak-to-peak amplitudes was not significantly different between the conventional and single channel EEG recordings (2.19 vs 2.27; P value was not significant).
|
|
1 seizure in a mean of 40.3% ± 16.8% of the 125 records with seizures (range: 21.6%–57.4%) and a mean of 25.5% ± 10.6% of the individual seizures (range: 12%–38%). Only 19 (15%) of 125 aEEG records had
1 seizure detected by all of the neonatologists, and only 1 (<1%) of 125 records had all of the seizures detected by every neonatologist. There were, however, no false-positive seizure detections by any neonatologist among the 19 control records. On multivariate analysis, factors related to correct seizure detection on aEEG included the neonatologists' experience with aEEG interpretation (comparing the 2 most experienced participants with all of the others combined), as well as seizure duration, frequency, and amplitude (Table 3). Of note, after adjusting for the interpreter's experience, neither EEG background characterization nor location of seizure onset was significantly associated with seizure detection by aEEG.
|
| DISCUSSION |
|---|
|
|
|---|
C4) is high, with 94% of all EEG records with seizures and 78% of all individual seizures detected in the raw EEG channel from which aEEG tracings are derived. However, in this study, the actual sensitivity of electrographic seizure detection by aEEG was low.
Five factors were significantly associated with correct seizure detection by aEEG: the neonatologists' level of expertise in aEEG interpretation, seizure appearance in the raw C3
C4 channel, seizure amplitude, seizure duration, and the number of seizures per hour. The neonatologists with the most aEEG experience identified 38% of all of the individual seizures and 57% of all of the seizure-containing records. However, the other neonatologists, each of whom had
1 year of aEEG experience, detected as few as 12% of individual seizures and 22% of seizure-containing records. Thus, even when interpreted by world experts, many individual electrographic seizures are inherently difficult to detect with aEEG.
A seizure's appearance in C3
C4 is intuitively related to its visibility on aEEG. It would not be possible to detect a seizure if it was not present in the raw EEG channel from which the aEEG was derived. A large proportion of individual seizures (78%) did appear in the C3
C4 channel, indicating that this location was well-selected. When this factor was included in the multivariate model, other variables describing seizure location (such as seizure onset in the frontal, central, temporal, or occipital electrodes) were not significantly associated with correct seizure detection.
The seizures' peak-to-peak amplitudes were significantly associated with their detection by aEEG. Because aEEG depends heavily on signal amplitude characteristics, detecting a seizure with this type of cerebral function monitor requires the seizure amplitude to be conspicuously higher than the interictal EEG background. The ratio of ictal to interictal peak-to-peak EEG amplitudes in our study was
2:1. For some seizures, this may prove to be too small to produce consistently discernible ictal deflections on the aEEG, especially if the events are also of brief duration.
Seizure duration was also significantly associated with correct seizure detection. The distribution of seizure duration is heavily weighted to brief events (Fig 3). aEEG is typically displayed at a paper speed of 6 cm/h. The average electrographic seizure detected in the raw C3
C4 channel (100 seconds) corresponds with a deflection of 1.6 mm on an aEEG trace. In isolation, this could be difficult to detect by visual inspection. However, the visual reinforcement of repetitive seizures made it easier to detect individual seizures in aEEG records with high seizure counts per hour, independent of seizure amplitude or duration.
|
In their study of neonatal encephalopathy, al Naqeeb et al13 described agreement on suspected seizures detected by aEEG among a neonatologist and 2 pediatrics residents (
= 0.76) on the aEEGs of 20 of 56 encephalopathic neonates. However, they did not have simultaneous conventional EEG to verify their seizure detections. Another study included 3 patients with electrographic seizures and found that seizures lasting <30 seconds were not detected by aEEG.20 Our results are consistent with this finding. The 2 most expert neonatologists in our study detected 22% of individual seizures lasting <30 seconds compared with 55% of seizures >30 seconds.
Toet et al15 reported a study of 33 simultaneously recorded 30-minute conventional EEG and aEEG traces. That study included 10 infants with seizures confirmed on conventional EEG, 8 of whom had
1 seizure detected by simultaneous aEEG. They also noted that low-amplitude seizures and those confined to the occipital or frontal electrodes were often missed on aEEG. In contrast, the current study shows that the location of seizure onset is not significantly associated with seizure detection on aEEG after adjustment for the seizure's appearance in C3
C4. However, we did confirm the finding that low-amplitude seizures were often overlooked on aEEG.
Rennie et al16 reported on nonexpert use of aEEG for neonatal seizure detection, using a sample of aEEG traces from 19 infants with seizures and 21 without. They found the sensitivity for individual seizure detection to be 38%–55% among their 4 interpreters, higher than our result of 12%–38%. Only 4 of their 19 patients with seizures on aEEG were identified by all of the neonatologists. The seizures in those records were relatively long (>1 minute) or of unspecified "high amplitude." The authors speculated that the poor sensitivity for neonatal seizure detection by aEEG was largely attributable to their interpreters' lack of experience. However, despite including both internationally recognized experts and less experienced clinicians in our study, we found similarly poor sensitivity. Only 19 of our 125 aEEG records were identified by all of the neonatologists as having seizures, and in only 1 of 125 did all of the interpreters recognize every seizure in the tracing. This implies that there are features inherent both to neonatal seizures and to the aEEG technique that limit the sensitivity of aEEG for neonatal seizure detection.
Several clinical studies of aEEG as a predictor of outcome after neonatal encephalopathy have commented on seizures. However, none had confirmation of the seizures by simultaneous conventional EEG. Toet et al21 found that the presence of seizures on aEEG at 3 hours of life predicted a poor outcome. However, the absence of seizures was not predictive of good outcome. This result could be partially because of low sensitivity of neonatal seizure detection on aEEG. Hellström-Westas et al14 found that repetitive seizures correlated with poor outcome after neonatal asphyxia. However, for 6 of these 7 infants, the seizures were clinical and were not detected on aEEG.
Our study had limitations. The EEG records were relatively brief (23–145 minutes) compared with the expanded recording times that are often clinically available with aEEG. It is likely that additional recording time could increase the percentage of records in which
1 seizure is identified, but this would not be expected to increase the proportion of individual seizures detected. Some interpreting neonatologists commented that the records were too brief for them to reach reliable conclusions about the presence of seizures. However, multicenter studies have used short aEEG traces in their protocols, including the CoolCap Trial,22 which used a screening 30-minute aEEG as an inclusion criterion. Because short-term aEEG monitoring is used in clinical practice and in research protocols, it was believed that the duration of our recordings was reasonable for the purpose of this study.
The neonatologists who interpreted our aEEG traces did not have access to the raw EEGs. Most modern aEEG equipment allows the interpreter to intermittently call up and display the raw EEG corresponding with a suspicious event observed on the compressed aEEG trace. Therefore, the participating neonatologists were instructed to mark as "seizure" the segments of the aEEG that were suspicious enough to trigger them to look at the raw EEG for confirmation.
The strengths of this study included the use of aEEG traces derived directly from conventional EEG (equivalent to simultaneously recording the 2 techniques), independent interpretation of conventional EEG and aEEG by experts in both methods, and the large sample size, which allowed for analysis of multiple variables of clinical interest (Table 3). As aEEG is introduced into more NICUs, the technique's strengths and shortcomings need to be investigated. Comprehensive studies, such as this contemporary characterization of neonatal seizures on conventional EEG, single-channel EEG, and aEEG, will help to clarify the appropriate role of aEEG in the monitoring and care of neonates with seizures.
| CONCLUSIONS |
|---|
|
|
|---|
| ACKNOWLEDGMENTS |
|---|
| FOOTNOTES |
|---|
Address correspondence to Renée A. Shellhaas, MD, Division of Neurology, 6th Floor, Wood Building, Children's Hospital of Philadelphia, 34th Street and Civic Center Boulevard, Philadelphia, PA 19104. E-mail: shellhaa{at}umich.edu
The authors have indicated they have no financial relationships relevant to this article to disclose.
These data were presented in abstract form at the annual meeting of the American Epilepsy Society; December 1–5, 2006; San Diego, CA.
| REFERENCES |
|---|
|
|
|---|
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||