a Department of Neonatology and Pediatric Intensive Care, University Childrens Hospital Greifswald, Greifswald, Germany
b Institute of Clinical Chemistry and Laboratory Medicine, University Hospital Greifswald, Greifswald, Germany
c Institute of Clinical Chemistry, Klinikum Harlaching, Munich Municipal Hospital, München, Germany
d Roche Diagnostics GmbH, Mannheim, Germany
| ABSTRACT |
|---|
|
|
|---|
METHODS. In a prospective study, bilirubin concentrations were determined with 9 different methods, ie, 3 skin test devices, 3 nonchemical photometric devices (including 2 blood gas analyzers), and 3 laboratory analyzers.
RESULTS. A total of 124 samples were obtained. All 3 laboratory methods showed very strong correlations with each other, and their means were used as comparison values. To these comparison values, the skin test devices had correlation coefficients between 0.961 and 0.966, and the nonchemical photometric devices between 0.980 and 0.994. Bland-Altman plots demonstrated good agreement with the comparison values for all nonchemical photometric devices. All skin test devices and 1 nonchemical photometric device underestimated bilirubin levels, particularly at high concentrations.
CONCLUSIONS. In the routine care of newborns, the first method for bilirubin testing should be a skin test. If the skin test result exceeds 200 µmol/L and other analytes are to be determined with a nonchemical photometric device, then bilirubin can be included in this analysis and the result trusted up to 250 µmol/L. If the skin test result exceeds 200 µmol/L and only bilirubin concentrations are needed, then a standard laboratory method is the first choice, to avoid repeated blood sampling. Bilirubin concentrations from nonchemical photometric devices that exceed 250 µmol/L should be confirmed with standard laboratory methods.
Key Words: bilirubin hyperbilirubinemia jaundice neonatal blood gas analysis infant newborn diseases
Abbreviations: TcBtranscutaneous bilirubin TBtotal bilirubin NISTNational Institute of Standards and Technology
JAUNDICE IS A common condition among neonates, caused by the combination of increased heme catabolism and physiologic immaturity of the liver in bilirubin conjugation and excretion.1 For most healthy term newborns, unconjugated bilirubin levels increase above 170 µmol/L (10 mg/dL) between the third and sixth days of life and decrease in the following days. Greater and more prolonged increases in bilirubin levels can be caused by hemolytic disorders (ABO or Rh factor incompatibility), glucose-6-phosphate dehydrogenase deficiency, or birth trauma.2 Clinically relevant hyperbilirubinemia is also seen among breastfed or premature newborns.3
High concentrations of bilirubin are toxic to the brain and may cause a specific type of bilirubin-related encephalopathy called kernicterus. This problem has been reemerging in recent years because of an increase in the number of healthy term newborns who receive less professional medical care during their first week of life, ie, those discharged early from the hospital or born at home.4 However, hyperbilirubinemia can be treated easily and with minimal adverse effects with phototherapy.5
Therefore, determination of bilirubin levels is performed for many newborns. The devices used for this purpose can essentially be grouped as (1) hand-held point-of-care devices for noninvasive reading of the skin to measure transcutaneous bilirubin (TcB) levels, (2) devices for nonchemical photometric measurement in blood samples (eg, blood gas analyzer), and (3) laboratory analyzers for photometric measurement of total bilirubin (TB) levels in serum or plasma after a chemical reaction.
A number of different methods are available for each of these 3 principal groups. Multiple investigations analyzing accuracy in bilirubin measurements with these different devices have been conducted. However, those studies usually compared only 2 or 3 devices. In addition, even routine laboratory methods for the determination of serum TB levels differ markedly.6 On the basis of these studies, it is difficult for physicians to decide which method should be used to determine bilirubin levels and, more importantly, how to interpret the results obtained. As a consequence, frequently hospitals develop their own guidelines, but these vary substantially. Therefore, we compared 9 frequently used methods for bilirubin determination for newborns, under routine conditions, to define their sequence of use.
| METHODS |
|---|
|
|
|---|
32 weeks and a birth weight of
1500 g. Infants receiving phototherapy before blood sampling were excluded. For each infant, bilirubin concentrations were determined with 9 different devices, ie, 3 skin test devices, 3 nonchemical photometric devices, and 3 laboratory analyzers. Skin testing and blood sampling for bilirubin determinations were performed simultaneously, at the time of routine metabolic screening or if there was a clinical indication for bilirubin determination. A maximum of 1 blood sample per day and a total of 3 on consecutive days could be obtained from each individual newborn. The study was conducted according to good clinical practice guidelines and was approved by the local ethics committee. Written informed consent was obtained from the parents.
Instruments
Skin Test Devices
Bilirubin measurements were performed at the lower end of the sternum. Bilirubin concentrations were estimated by measuring the optical density of the skin at certain wavelengths. In our study, we used 3 TcB devices. To reduce variability, only 1 instrument per device type was used throughout the study period. The Biliblitz JM-102 device (Minolta, Bad Ems, Germany) gives its result as an arbitrary number; the instrument must be calibrated by the user. In our study, we retrospectively transformed the arbitrary numbers (range: 6.924.6; mean: 15.9) into bilirubin concentrations by using our comparison (Passing-Bablok regression analysis; TB [µmol/L] = 20.2 · [JM-102 reading] 167.02). The Biliblitz JM-103 device (Minolta) represents an improvement by presenting the bilirubin result in micromoles per liter or milligrams per deciliter. With the BiliCheck analyzer (Medela, Eching, Germany), 5 replicate measurements must be performed to obtain a single result, given as the mean bilirubin concentration in micromoles per liter or milligrams per deciliter.
Nonchemical Photometric Devices
Twin Beam (Ginevri, Rome, Italy), ABL 735 (Radiometer, Copenhagen, Denmark), and Roche OMNI S (Roche Diagnostics, Graz, Austria) analyzers were developed to measure TB levels, with minimal blood volumes, at the point of patient care. The Twin Beam analyzer (Twin Beam Plus version) measures bilirubin levels in plasma at 455 nm and 575 nm. This necessitates centrifugation of the blood sample. The ABL 735 analyzer (software version V3.811) is a blood gas analyzer with a cooximetry module. Bilirubin levels are measured spectrophotometrically at 128 reading points between 478 nm and 672 nm, in addition to the other cooximetry parameters. The Roche OMNI S analyzer (software version 1.0131) is also a blood gas analyzer with cooximetry. Bilirubin levels are measured at 512 reading points between 460 nm and 660 nm. The ABL 735 and Roche OMNI S analyzers both hemolyze the blood sample ultrasonically within the cooximetry module before measurement.
Laboratory Analyzers
Hitachi 912 (Roche Diagnostics), Dimension RxL (Dade Behring, Schwalbach, Germany), and Vitros 250 (Ortho Clinical Diagnostics, Neckargemünd, Germany) analyzers are clinical chemical analyzers used in core laboratories. The methods for determination of TB levels for the Hitachi 912 and Dimension RxL analyzers are diazo methods; the method for the Vitros analyzer is a direct spectrophotometric assay.
Results obtained from measurements of patients samples were used only if quality control results for the device on that day were in the declared range. Table 1 presents the principles of bilirubin measurement for all devices used.
|
The Vitros 250 was available on site only until September 2003. Therefore, in addition to the 29 determinations performed up to that time, bilirubin levels were measured for all samples with the identical instrument at the hospital in Prerow, Germany, in April 2004. The determinations were performed with remaining plasma that had been stored at 70°C. The difference between measurements in Greifswald and in Prerow with the Vitros 250 was within the range observed for the 3 central laboratory methods (y = 0.967x 0.367 µmol/L; n = 29).
In our study, we used premarketing software on the Roche OMNI S analyzer (version 1.01-31). With the marketed software, results are given only for whole-blood samples with bilirubin concentrations of >51 µmol/L (>3 mg/dL).
Analytical Performance
To investigate accuracy, quality controls were prepared from certified reference material SRM 916a (98.3 ± 0.3% unconjugated bilirubin), obtained from the National Institute of Standards and Technology (NIST) (Gaithersburg, MD). Quality controls were prepared at 4 different levels (NIST levels of 0, 34, 170, and 340 µmol/L) in modified Tris buffer or in human fresh frozen plasma, as described,7 and were immediately stored at 70°C until needed. These quality controls were also used to assess within-run imprecision (21 repetitions each at 0, 34, 170, and 340 µmol/L) and day-to-day imprecision (21 days for each of the 4 levels), according to National Committee for Clinical Laboratory Standards guidelines. In addition, NIST reference material in plasma was mixed with washed erythrocytes from umbilical cord blood and used for within-run (4 repetitions for each level with all devices) and day-to-day (16 days for each level with ABL 735 and Roche OMNI S) imprecision assessments. Linearity assessments were performed with duplicate measurements for samples at 13 levels from 0 to 340 µmol/L. In addition to these measurements, daily quality control procedures were performed for all instruments according to manufacturers instructions.
Statistical Analyses
Data were analyzed with Microsoft Excel 2000 (Microsoft, Redmond, WA) and MedCalc (version 4.31.010) software (MedCalc Software, Mariakerke, Belgium), with Passing-Bablok regression analyses, Bland-Altman plots, and receiver operating characteristic curves.8
| RESULTS |
|---|
|
|
|---|
The 3 laboratory methods met the expectations with respect to imprecision (<5%) and accuracy (<5%) with quality control material, as described above. In addition, we used 1 control serum sample from an independent manufacturer for the 3 laboratory analyzers, ie, Hitachi 912, Dimension RxL, and Vitros 250, with declared target values of 358.5, 377.5, and 335.0 µmol/L, respectively. For these 3 instruments, the means of the daily serum bilirubin determinations were 341.3 µmol/L (Hitachi 912), 349.3 µmol/L (Dimension RxL), and 337.2 µmol/L (Vitros 250).
The measurements of patient samples with the 3 standard laboratory analyzers correlated strongly with each other (Hitachi 912 versus Dimension RxL: y = 1.019x + 3.254 µmol/L; r = 0.994; Dimension RxL versus Vitros 250: y = 0.943x 2.629 µmol/L; r = 0.994; Vitros 250 versus Hitachi 912: y = 1.045x 1.516 µmol/L; r = 0.993 µmol/L). Therefore, and because no standard test for bilirubin determination is available, we used the mean of the Hitachi 912, Dimension RxL, and Vitros 250 measurements for comparisons with the other 6 devices. This mean is referred to as our comparison value. Results from all 3 instruments were available for all except 9 samples, for which only 2 measurements were performed.
Passing-Bablok regression analyses gave correlation coefficients for the comparison of 0.961 to 0.966 for the 3 skin test devices, of which the BiliCheck had the best result. Higher correlation coefficients were seen for all 3 nonchemical photometric devices (r = 0.9800.994); of these, the Twin Beam had the best correlation coefficient (Table 2).
|
|
222 µmol/L (
13 mg/dL). Among the nonchemical photometric devices, the Roche OMNI S analyzer had the highest bilirubin value at both concentrations.
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
In 1980, Minolta/Air Shields presented a jaundice meter for noninvasive bilirubin measurements.10 This device and its successor (JM-102) were studied extensively1115 and were found to have 2 major disadvantages, ie, the instrument must be calibrated by the user to convert the results given as arbitrary numbers into serum bilirubin concentrations, and readings are affected by the race and age of the infant.16
The JM-103 and the BiliCheck are improved devices for measuring TcB levels without an influence of skin pigmentation.14,15,1719 Several studies demonstrated that these skin test devices were useful in screening neonates for hyperbilirubinemia but could not replace TB measurements because of insufficient accuracy in predicting serum bilirubin concentrations.13,2023 However, with the use of TcB devices, blood sampling can be avoided for 66 to 80 of 100 infants.22,24,25
In our study, all 3 skin test devices showed good correlation with our comparison values, with the BiliCheck having the best agreement. According to the literature, skin test devices tend to underestimate bilirubin concentrations.15,1821,23,24 This underestimation is pronounced at high bilirubin concentrations.20,23 Mean differences reported were up to 11.6 µmol/L for the JM-102, up to 3.3 µmol/L for the JM-103, and up to 30 µmol/L for the BiliCheck.15,18,19,21,24 Our study found identical underestimations for the JM-102 and JM-103 but a slight overestimation by the BiliCheck (on average, +8.3 µmol/L). At the clinically most relevant high bilirubin concentrations above 200 to 250 µmol/L (11.714.6 mg/dL), all skin test devices underestimated levels, with maximal differences between 39 and 58 µmol/L (between 2.3 and 3.4 mg/dL). These results mark the limitations of these devices.
For physicians caring for neonates, it is essential to know the TcB values up to which they can trust the skin test device and avoid TB measurements without missing an infant in need of therapeutic intervention. Therefore, for clinical purposes, it is important to define a cutoff value describing the highest measurement result at which each device identifies correctly all infants with levels above a defined bilirubin concentration. For 257 µmol/L (15 mg/dL) TB, according to our comparison, these cutoff values were between 209 and 224 µmol/L (between 12.2 and 13.1 mg/dL) for the different devices. Results up to these values need not be confirmed with a laboratory test. Among the skin test devices, the JM-102 had the highest cutoff value. If no blood tests were performed below this cutoff value, then 93% of blood sampling could have been avoided for our study population. The BiliCheck had an almost identical cutoff value and the JM-103 a lower one, up to which these devices identified correctly all infants with levels of >257 µmol/L (>15 mg/dL). With those cutoff values, again 93% of blood sampling could have been avoided.
With the development of blood gas analyzers, near-patient bilirubin determinations from blood samples became available. These devices are popular because multiple other clinically important analytes, such as pH, sodium levels, and calcium levels, can be determined from the same blood sample. Results are provided quickly, and the amount of blood needed is small.26 Transport to the laboratory is no longer necessary. However, costs are usually higher and quality assurance is more difficult to control than with laboratory methods.
In our study, we used 2 different blood gas analyzers, the ABL 735 analyzer and the newly developed Roche OMNI S analyzer, which measure bilirubin levels through multiple-wavelength photometry. The Twin Beam is a bilirubinometer that measures bilirubin levels quickly and at almost no cost. As a disadvantage of this device, the sample must be centrifuged before measurement of bilirubin levels; if this is performed in a capillary tube, however, then both hematocrit and bilirubin can be measured from the same sample.
Results from all 3 nonchemical photometric devices correlated well with TB values. The Twin Beam had the highest correlation coefficient, but the device tended to underestimate bilirubin levels, particularly at high concentrations. This underestimation was especially evident in all imprecision measurements. In contrast to other authors who reported correlation coefficients of 0.76 and 0.95 for ABL 735 measurements versus TB values,27,28 we found a higher correlation coefficient (0.987). This could be attributable to newer software implemented since those studies were performed. These software changes might also be the cause of the underestimation by, on average, 17 µmol/L (1 mg/dL) seen in our study, compared with 3.5 µmol/L (0.2 mg/dL) found previously.27
For the 3 nonchemical photometric devices also, we determined cutoff values above which blood samples should be sent to the laboratory in order not to miss an infant in need of therapeutic intervention. In our study, the Roche OMNI S analyzer had the highest cutoff value up to which all neonates were identified correctly. If no blood samples up to this cutoff value had been sent to the laboratory, then 95% of blood sample transports could have been avoided for our study population. The ABL 735 and the Twin Beam identified correctly all infants at lower cutoff values. With those cutoff values, 92% of blood sample transports could have been avoided.
Results from all 3 laboratory methods used in our study had extremely strong correlations with each other. Therefore, their mean was considered to be close to the "true" bilirubin concentration and served as a comparison value, because a true standard test for bilirubin determination has not been established. Only a proposed reference method has been published; it is neither used frequently nor accepted widely, because of its requirement for a manual procedure.7
Subjects in our study were white term infants. Also, bilirubin measurements were performed on the third day of life, on average. Therefore, our conclusions should be used with care for low birth weight or preterm infants, and those concerning the skin test devices might not be appropriate for nonwhite or older infants. Only 7% of our samples had bilirubin concentrations of >257 µmol/L (>15 mg/dL). Additional research should be conducted, particularly for patients with high bilirubin levels.
| CONCLUSIONS |
|---|
|
|
|---|
Nonchemical photometric devices give more-accurate information on bilirubin concentrations than do skin test devices, but a blood sample is necessary. Among the nonchemical photometric devices, the Twin Beam presented the best results. However, readings above 250 µmol/L should be interpreted with care, because of the observed underestimation at high concentrations. Although the analysis itself is inexpensive with the Twin Beam, the requirement for sample preparation makes blood gas analyzers a more convenient choice. Their readings can be trusted up to 250 µmol/L, and they are the first choice, especially if blood gas analysis is indicated for other reasons. If skin test results exceed 200 µmol/L and only bilirubin concentrations are needed, then standard laboratory methods are the first choice, to avoid repeated blood sampling. Bilirubin concentration results from blood gas analyzers that exceed 250 µmol/L should be confirmed with standard laboratory methods.
| ACKNOWLEDGMENTS |
|---|
We thank Dade Behring (Schwalbach, Germany), Ginevri (Rome, Italy), Medela (Eching, Germany), Minolta (Bad Ems, Germany), Ortho Clinical Diagnostics (Neckargemünd, Germany), and Radiometer (Copenhagen, Denmark) for providing their devices and excellent help. We are in debt to all infants and their parents who took part in this study, as well as to all nurses, doctors, and technicians for their help and support.
| FOOTNOTES |
|---|
Address correspondence to Helmut Küster, MD, Department of Neonatology and Pediatric Intensive Care, University Childrens Hospital, Soldmannstrasse 15, D-17475 Greifswald, Germany. E-mail: kuester{at}uni-greifswald.de
Financial Disclosure: Dr Goerlach-Graw is employed by Roche Diagnostics GmbH (Mannheim, Germany).
| REFERENCES |
|---|
|
|
|---|