ELECTRONIC ARTICLE |


* Department of Pediatrics, University of Padova, Padova, Italy
Primary Care Pediatrics, Federazione Italiana Medici Pediatri, Italy
Valeas SpA, Milano, Italy
| ABSTRACT |
|---|
|
|
|---|
Methods. Ten primary care pediatricians undertook a spirometry training program that was led by 2 pediatric pulmonologists from the Pediatric Department of the University of Padova. After the pediatricians' training, children with asthma or persistent cough underwent a spirometric test in the pediatrician's office and at a pulmonary function (PF) laboratory, in the same day in random order. Both spirometric tests were performed with a portable turbine flow sensor spirometer. We assessed the quality of the spirometric tests and compared a range of PF parameters obtained in the pediatricians' offices and in the PF laboratory according to the Bland and Altman method.
Results. A total of 109 children (mean age: 10.4 years; range: 615) were included in the study. Eighty-five (78%) of the spirometric tests that were performed in the pediatricians' offices met all of the acceptability and reproducibility criteria. The 24 unacceptable test results were attributable largely to a slow start and failure to satisfy end-of-test criteria. Only the 85 acceptable spirometric tests were considered for analysis. The agreement between the spirometric tests that were performed in the pediatrician's office and in the PF laboratory was good for the key parameters (forced vital capacity, forced expiratory volume in 1 second, and forced expiratory flow between 25% and 75%). The repeatability coefficient was 0.26 L for forced expiratory volume in 1 second (83 of 85 values fall within this range), 0.30 L for forced vital capacity (81 values fall within this range), and 0.58 L/s for forced expiratory flow between 25% and 75% (82 values fall within this range). In 79% of cases, the primary care pediatricians interpreted the spirometric tests correctly.
Conclusions. It seems justifiable to perform spirometry in pediatric primary care, but an integrated approach involving both the primary care pediatrician and certified pediatric respiratory medicine centers is recommended because effective training and quality assurance are vital prerequisites for successful spirometry.
Key Words: office spirometry primary care pediatrics asthma children
Abbreviations: PF, pulmonary function FEV1, forced expiratory volume in 1 second ATS, American Thoracic Society FVC, forced vital capacity FET, forced expiratory time FEF2575, forced expiratory flow between 25% and 75% of expired FVC ICC, intraclass correlation coefficient GP, general practitioner
Pulmonary function (PF) tests are useful in both the diagnosis and the monitoring of lung disease. The international guidelines for asthma management have endorsed the use of objective lung function measures for assigning a severity rating to patients with asthma and guide asthma therapy.1 Although children's and their parents' reporting of asthma symptoms is important in staging and managing pediatric asthma, many children and parents do not perceive asthma symptoms adequately, and children with long-standing airway obstruction are less likely to report dyspnea than children with acute-onset airway obstruction.24 In addition, physical findings seem to be inadequate for assessing obstruction that may be present despite a normal physical examination.5
Despite a large body of evidence showing that airway obstruction in children with asthma is associated with ongoing respiratory morbidity6,7 and a reduced forced expiratory volume in 1 second (FEV1) in adulthood,810 spirometry is not routinely performed by physicians who treat children with asthma as an objective measure of airway obstruction. According to the Asthma Insights and Reality Europe Study (a survey that assesses the current level of asthma control in Western Europe), a large proportion of children with asthma are treated without lung function measurements, and physicians base their treatment decisions on symptom reports and auscultation.11
Ideally, spirometry should be available on-site in primary care practices. However, despite the increasing use of spirometric tests in primary health care, little is known about their validity.1217 Spirometry is often regarded as a simple, noninvasive screening test, but careful consideration clearly needs to be given to a number of aspects, including optimal test performance, adherence to standard acceptability and repeatability criteria, and accurate, well-informed interpretation of the results.
Portable, lightweight spirometry instruments that can be wired into a computer are now available, making spirometry technically feasible at the primary care level. Advantages of the newly proposed category of office spirometers include lower instrument cost, small size, and an improved quality-assurance program.16,18
No previous studies have formally assessed spirometry performance in primary care pediatric practice. We aimed to determine the quality of the spirometric tests that are performed in primary care pediatric practice after formal training for the pediatrician was provided. For this purpose, spirometric tests that were performed by primary care pediatricians were assessed for acceptability and reproducibility, and the resulting spirometric data were compared with those obtained at a PF laboratory using the same portable spirometer. The primary care pediatrician's interpretation of the spirometric data were also assessed. Finally, to validate the portable spirometer, the data that it produced in the laboratory were compared with those that were obtained with a gold-standard laboratory-based volume spirometer.
| METHODS |
|---|
|
|
|---|
Children who had a diagnosis of asthma1 or cough persisting for >15 days and were aged 6 to 15 years were selected for the study (Table 1). They were recruited during routine visits to the pediatricians' offices. All parents gave their written informed consent.
|
Spirometry Tests
Data collection started after the second training session and lasted for 1 month. The pediatricians' offices and laboratory all were equipped with the same portable spirometer (Microloop) and spirometry computer software (Spida 5). The spirometer uses a turbine flow sensor. The unit is provided with built-in quality assurance features, based on American Thoracic Society (ATS) criteria.21 The spirometry software displays real-time flow-volume and volume-time curves and produces built-in quality assurance prompts, which are displayed after each unacceptable blow (eg, slow start, abrupt end, short blow) and to indicate the variability between the 2 highest FEV1 and forced vital capacity (FVC) values. Spirometers were checked for reading errors by a technician before the study began, using a certified 3-L syringe. All spirometers had a deviation of <3% in the volume reading.
Patients were instructed to abstain from short-acting bronchodilators for 8 hours and long-acting bronchodilators for 12 hours before taking the test. Each patient underwent 2 spirometric tests (forced expiratory maneuvers) in random order in the same day. One test was performed at the pediatrician's office; the other was performed at the PF laboratory by 1 of the pulmonologists. Both the pediatricians and the pulmonologists were unaware of the results of the other spirometric test. All spirometric tests were conducted with the patient standing and wearing a nose clip. Patients were instructed to inhale completely before inserting the mouthpiece while holding their breath, then exhale forcefully into the spirometer for as long as they could. At least 3 acceptable and 2 reproducible forced expiratory maneuvers were required for each test. Table 2 shows the acceptability and reproducibility criteria. According to ATS recommendations, a forced expiratory time (FET) of <6 seconds is acceptable in children,21 and we considered a FET
2 seconds as acceptable.22 The expiratory spirometric indices (including FEV1 and FVC) of the maneuver with the highest sum of FEV1 + FVC were saved and used for analysis. Flow-volume and volume-time curves and spirometric parameters results were printed.
|
Outcomes and Statistical Analysis
Acceptability Criteria
Individual spirograms were judged to be "acceptable" when all of the following requirements were met: good start (as defined by a back extrapolated volume <5% of FVC), satisfactory exhalation for
2 seconds, no abrupt end, and no cough during the maneuver (Table 2).
Reproducibility Criteria
After 3 acceptable spirograms had been identified, the test was judged to be reproducible when the two highest FVC and FEV1 values were within 5% of each other.
Comparison Between Pediatricians' Offices and Laboratory Spirometric Tests With the Portable Spirometer
One of the primary outcomes was the within-subject difference between laboratory and primary care office spirometric test results in terms of
FEV1,
FVC, and forced expiratory flow between 25% and 75% of expired FVC (
FEF2575).
Interpretation
For each spirometric test result, the primary practice pediatrician was asked to record an interpretation, choosing from the following options: normal, mild obstruction, moderate obstruction, moderate to severe obstruction, severe obstruction, and restriction.23 These records then were reviewed by the 2 pediatric pulmonologists, who marked the primary care pediatrician's interpretations as "correct" or "incorrect."
Statistical Analysis
For the statistical analysis, we performed a power calculation that showed that we needed 85 patients per spirometry group to have 80.3% power to detect a 7% or greater difference in spirometric parameters between spirometry performed in the pediatric office and in the PF laboratory (with
set at 5%). FVC, FEV1, and FEF2575 values are expressed as median and interquartile range. FVC, FEV1, and FEF2575 of the spirometry tests that were performed with the portable spirometers in the pediatricians' offices and the PF laboratory were compared using an intraclass correlation coefficient (ICC). In addition, for assessing the agreement between the 2 spirometric test results, crude mean
FVC,
FEV1, and
FEF2575 were calculated by subtracting a patient's laboratory value from the pediatric practice value. Mean values for these parameters with 95% confidence interval were calculated, and difference versus mean plots and accompanying limits of agreement (±1.96 times the SD of the differences) were produced to express the variability between laboratory and pediatric practice measurements according to the Bland and Altman method.24 The Bland and Altman method was used to assess the agreement between the spirometric parameters that were obtained in the PF laboratory with the portable and the laboratory-based volume spirometers.
| RESULTS |
|---|
|
|
|---|
Spirometric Test Quality
All of the spirometric test results were checked by the 2 pediatric pulmonologists. The analysis of the tests that were performed by the pediatricians in primary care practice revealed that 85 (78%) of 109 tests met all of the acceptability and reproducibility criteria (at least 3 acceptable and 2 reproducible curves). Most of the unacceptable blows were so judged on the grounds of an inadequate duration of exhalation or slow start (Table 3). In 6 cases, the spirometric test results were not reproducible. Twelve of the 24 unacceptable spirometry tests had been performed by children who were
8 years of age. In 11 of 24 cases, the children were performing their first spirometry.
|
FEV1,
FVC, and
FEF2575 values did not vary systematically over the range of measurements.
|
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
Laboratory-based spirometry is the "gold standard" for assessing lung function in children with asthma, but the costs of the technician and equipment led to peak flowmeters' often being used instead for office and home monitoring of lung function in children. Although peak expiratory flow monitoring and asthma diaries are useful tools in selected patients, there are major limitations to their widespread use in children with asthma.25,26 Portable spirometers are now available with the capability for collecting a wide range of PF parameters that are clinically and epidemiologically more useful than peak expiratory flow, because they are more sensitive to functional status changes in asthma.16,18
The validity of spirometric tests is a prerequisite for their use as a tool for the diagnosis, monitoring, and management of respiratory diseases. Little is known about the validity of spirometric tests in adult patients in the primary care setting, however, and it has been reported that at least one third of the tests that are performed by the general practitioner (GP) fail to meet the quality criteria that are applicable to PF laboratories.20,27
The validity of spirometric tests that are performed in general practice was investigated recently by Schermer et al15 by comparing them with the tests that are performed in patients with chronic obstructive pulmonary disease using the same spirometer at a PF laboratory. After a spirometry training program, the authors found the validity and the quality of the spirometric tests that were conducted by the GPs satisfactory, but the consistency between the laboratory's and the GP's measurements seemed to be limited. No studies have hitherto formally addressed spirometry performance in primary care pediatric practice.
Although seemingly straightforward, spirometry is an effort-dependent test that requires a cooperative patient and a trained operator to administer the test. We used a spirometer with built-in quality assurance prompts on the assumption that this would improve the quality of the spirometry, but 22% of the tests nonetheless failed to meet the acceptability and reproducibility criteria. Most of the errors were attributable to slow start or short blow.
In our study, the children were instructed to exhale for as long as they could, and a FET
2 seconds was considered valid. The ATS standards recommend an expiratory time of
6 seconds in adult patients,21 but, recognizing that children have difficulty in meeting such end-of-test criteria, they say that shorter exhalation times are acceptable in children. Assuming that the flow-volume curve shows no abrupt termination of expiratory flow and that the volume/time curve shows a plateau, it has been suggested that the minimum FET should be reduced to 2 seconds for children who are
8 years and to 1 second for children who are younger than 8 years.22 Other researchers suggested taking 3 seconds28 or 4 seconds29 as the target exhalation times.
A previous study in adult patients with chronic obstructive pulmonary disease demonstrated a slight but significant difference between the spirometric indexes that were measured by the GP's staff and those that were measured at a PF laboratory, suggesting that an interchangeable use of these measurements should be avoided.15 We found no difference in FVC, FEV1, and FEF2575 between the pediatrician's and the PF laboratory's results of tests that were performed with the portable spirometer. In addition, we demonstrated that the portable spirometer that we used provides an accurate and reliable measure of PF when compared with a gold-standard laboratory-based device, as shown by the consistency that emerged between the 2 spirometers across the key spirometric parameters.
Although we primarily addressed the quality of the test's performance, the pediatricians' interpretation of the results was incorrect in 21% of the cases reviewed. In many cases, this misinterpretation involved the failure to detect or an underestimation of airway obstruction, so any clinical decisions that are based on these results would compound, rather than reduce, the underdiagnosis and undertreatment of asthma. Accurate interpretation relies not only on a well-performed procedure but also on an appreciation of physiology and a thorough clinical knowledge of the patient. Of course, longer, more intensive training might have produced better results in terms of the interpretation of the findings.
Proper training for the technician/physician who performs the test is perhaps the most important factor in ensuring good-quality spirometric testing.20,30 As stated in 1991 by the ATS,23 the largest single source of within-subject variability is improper performance of spirometry; therefore, effective training and quality assurance are vital prerequisites for successful spirometry. After adequate training, it is also important to have continual competence assessments and to review the test results carefully. It has been demonstrated that, even in a dedicated research setting paying meticulous attention to quality, a technician's performance declines over time, but the quality of spirometry can improve dramatically and be maintained by routinely monitoring test session quality and prompting individual feedback.31,32
In both the United States and Europe, there are accredited training centers for pediatric respiratory medicine with modules on PF testing. Because effective training and quality assurance are vital prerequisites for successful spirometry, we believe that only approved training centers for pediatric respiratory medicine should certify that a primary care pediatrician has acquired the competence for performing and interpreting a spirometric test.
In conclusion, with sufficient training for the physicians and their staff, it seems justifiable to perform spirometry in the pediatric primary care setting. This could improve asthma monitoring in children, as recommended by the international guidelines.1 If spirometry is to be promoted as a screening tool in primary care practice, however, then it is important to pay careful attention to ensuring that quality standards are met. To achieve this goal, an integrated approach is recommended, involving both the primary care pediatrician and the certified pediatric respiratory medicine centers.
| ACKNOWLEDGMENTS |
|---|
Members of the Working Group: Franco Balliana, MD
; Eugenio Baraldi, MD*; Gianfranco Battaglini, MD
; Raffaele Braga
; Armando Cirillo, MD
; Annamaria De Marchi, MD
; Mattia Doria, MD
; Mario Fama, MD
; Giorgio Meneghelli, MD
; Angela Pasinato, MD
; Andrea Passarella, MD
; Flavio Semenzato, MD
; and Stefania Zanconato, MD, PhD*.
We thank Dr Silvia Carraro for valued help with statistical analysis.
| FOOTNOTES |
|---|
Address correspondence to Stefania Zanconato, MD, PhD, Department of Pediatrics, University of Padova, Via Giustiniani 3, 35128 Padova, Italy. E-mail: zanconato{at}pediatria.unipd.it
No conflict of interest declared.