PEDIATRICS Vol. 108 No. 3 September 2001, pp. 682-685
Work of Breathing During Constant- and Variable-Flow Nasal Continuous Positive Airway Pressure in Preterm Neonates
,
, §,
, §,
, §, and
From the * University of Medicine and Dentistry of New
Jersey/Robert Wood Johnson Medical School, Camden, New Jersey;
Background. Constant-flow nasal
continuous positive airway pressure (NCPAP) often is used in preterm
neonates to recruit and maintain lung volume. Physical model studies
indicate that a variable-flow NCPAP device provides more stable volume
recruitment with less imposed work of breathing (WOB). Although
superior lung recruitment with variable-flow NCPAP has been
demonstrated in preterm neonates, corroborating WOB data are lacking.
Objective. To measure and compare WOB associated with the
use of variable-flow versus constant-flow NCPAP in preterm
neonates.
Methods. Twenty-four preterm infants who were receiving
constant-flow NCPAP (means, SD) and had birth weight of 1024 ± 253 g, gestational age of 28 ± 1.7 weeks, age of 14 ± 13 days, and FIO2 of 0.3 ± 0.1 were
studied. Variable-flow and constant-flow NCPAP were applied in random
order. We measured changes in lung volume and tidal ventilation
(VT) by DC-coupled/calibrated respiratory inductance plethysmography as well as esophageal pressures at NCPAP of 8, 6, 4, and 0 cm H2O. Inspiratory WOB (WOBI) and lung
compliance were calculated from the esophageal pressure and
VT data using standard methods. WOB was divided by
VT to standardize the results.
Results. WOBI decreased at all CPAP levels
with variable-flow NCPAP, with a maximal decrease at 4 cm
H2O. WOBI increased at all CPAP levels with
constant-flow CPAP. Lung compliance increased at all NCPAP levels with
variable-flow, with a relative decrease at 8 cm H2O,
whereas it increased only at 8 cm H2O with
constant-flow NCPAP. Compared with constant-flow NCPAP,
WOBI was 13% to 29% lower with variable-flow NCPAP.
Conclusion. WOBI is decreased with
variable-flow NCPAP compared with constant-flow NCPAP. The increase in
WOBI with constant-flow NCPAP indicates the presence of
appreciable imposed WOB with this device. Our study, performed in
neonates with little lung disease, indicates the possibility of lung
overdistention at CPAP of 6 to 8 cm H2O with the
variable-flow device. Further study is necessary to determine the
efficacy of variable-flow NCPAP in neonates with significant lung
disease and its use over extended periods of time.continuous-flow and variable-flow NCPAP, work of breathing, premature
neonates, lung compliance.
Virtua-West Jersey Hospital, Voorhees, New Jersey; § Department of
Pediatrics, Division of Neonatology, The Children's Regional Hospital
at Cooper Hospital/UMC, Camden, New Jersey; and
Mercy Children's
Hospital at St. Vincent Mercy Medical Center and Department of
Pediatrics, Medical College of Ohio, Toledo, Ohio.
![]()
ABSTRACT
Top
Abstract
Methods
Results
Discussion
Conclusion
References
Nasal continuous positive airway pressure (NCPAP)
often is used in preterm neonates to recruit and maintain lung volume
(VL). NCPAP usually is provided by varying the
resistance to exhalation while constant gas flow is delivered by a
neonatal ventilator through nasal prongs (constant-flow NCPAP).
An NCPAP device that uses demand or variable gas flow (variable-flow
NCPAP) is available.1,2 Physical model studies comparing
constant-flow and variable-flow NCPAP report a relative decrease in
airway pressure variability during breathing with variable-flow NCPAP.
This indicates a potential for superior lung recruitment and
maintenance of VL.1,2 Moreover, Klausner et al2 found that the imposed work of breathing (WOB) with the variable-flow NCPAP prongs was one fourth that of
conventional constant-flow NCPAP prongs, ascribing this to differences
in prong design. Infant data to support these findings are scarce. In
particular, we are unaware of any published data on WOB during
variable-flow NCPAP support.
We recently demonstrated that variable-flow NCPAP provides superior
lung recruitment for similar CPAP levels in preterm
neonates.3 However, the gas flows required to generate an
equivalent CPAP with the variable-flow device generally are greater
than with the constant-flow device. Although the design of the
variable-flow NCPAP prongs allows for excess gas flow to be diverted
away from the patient, it is unclear whether WOB is affected. Thus, our objective was to measure and compare WOB associated with the use of
variable-flow versus constant-flow NCPAP in preterm neonates.
This study was approved by the Institutional Review Board of
Cooper Hospital/University Medical Center. Premature infants who
weighed <1800 g at birth and were receiving constant-flow NCPAP for
apnea or mild respiratory distress were eligible for enrollment,
provided that they were otherwise medically stable. Informed parental
consent was obtained before testing.
Data Acquisition and Analysis
The study design and data acquisition have been described in
detail elsewhere.3 Briefly, NCPAP was provided to each infant with each of 3 devices applied in random order: 1) a modified nasal canula, 2) a standard nasal CPAP prongs (constant-flow), and 3)
the new variable-flow NCPAP generator with prongs (variable-flow). Because a nasal canula is not used commonly to provide NCPAP, our WOB
analysis was restricted to the 2 NCPAP delivery systems using
prongs:
![]()
METHODS
Top
Abstract
Methods
Results
Discussion
Conclusion
References
Measurements were performed after a feeding to facilitate quiet sleep. Sedation was not used in any infant. Infants were instrumented and placed in the supine position. Instrumentation consisted of the following: 1) respiratory inductance plethysmography (RIP) bands that were fitted around the chest and the abdomen (Respiband Plus, SensorMedics Corp, Yorba Linda, CA; and Nims Inc, Miami Beach, FL); 2) insertion of a neonatal esophageal balloon catheter (Ackrad Laboratories, Cranford, NJ) at the level of the lower third of the trachea to estimate intrapleural pressure from the esophageal pressure (Pes); and 3) placement of a thermistor (BreathSensor, Nellcor Puritan Bennett, Eden Prairie, MN) to detect and continuously record (EdenTrace II Plus, EdenTec, Eden Prairie, MN) air leaks from the mouth.
RIP was used to measure lung volume changes
(
VL) and tidal ventilation
(VT) from its direct current (DC) and alternating current components, respectively (Somnostar, SensorMedics Corp., Yorba
Linda, CA).4,5 After the RIP bands were placed, RIP was
calibrated by direct comparison6,7 to leak-free flow and
volume data measured by face mask pneumotachography (Neonatal Flow
Sensor #7218 [dead space 0.8 mL], Novametrix, Wallingford, CT).
Before measurements began, proper placement of the Pes balloon catheter
was checked by continuous on-line monitoring of Pes and adjusted until
a high correlation (r2 > 0.90) was obtained between airway opening pressure and Pes during spontaneous breathing by
the infant against an occluded airway.8 When necessary, the infant's mouth was closed gently during data collection to stop
any air leak, and data with air leak at the mouth were not used. Airway
flow and Pes from a series of 10 to 15 leak-free breaths were
integrated to calculate VT. All signals were
sampled at 100 Hz, monitored on-line, and stored on a computer for
later analysis. Patients initially were placed on NCPAP of 8 cm
H2O to allow similar lung recruitment. With each
device,
VL, VT, and Pes
were measured at NCPAP of 8, 6, 4, and 0 cm H2O.
Infants were kept for 3 to 5 minutes at each CPAP level. The leak-free breaths spanning the last 20 to 30 seconds at each setting were selected for subsequent analysis.
Lung compliance (CL) was calculated from VT and Pes data using standard methods.8 Inspiratory WOB (WOBI) was calculated from the area subtended by the inspiratory limb of the Pes-VT curve, according to Campbell's diagram.9 Resistive WOB (RWOB) was calculated as the area between the inspiratory and expiratory loops of the Pes-VT curve. For standardizing the results from different infants and for varying breathing amplitudes, WOB was divided by VT.
Sample Size Calculation/Statistical Analysis
Sample size was based on finding a clinically significant
VL between any 2 NCPAP devices as previously
reported by us.3 A sample size of 28 to 32 patients is
required to attain a statistical power of 0.8 to 0.85 when one assumes
a 20% difference to be significant with a 0.05 significance
criterion.10 Dependent variables were analyzed as mixed
linear models in a randomized block factorial design.11
Devices and CPAP levels were considered fixed effects, and patients
were treated as random blocks. Differences between devices, within
NCPAP levels, and devices-within-NCPAP levels were tested using least
square, pair-wise mean comparisons. The effects of missing cells were
adjusted for by using Satterthwaite approximations.11
P < .05 was considered to be significant.
| |
RESULTS |
|---|
|
|
|---|
Thirty-five infants were recruited for study. Of these, Pes monitoring and RIP data that allowed accurate calculations of WOB were obtained in 24 infants. The patient characteristics of these 24 infants including baseline CL are shown in Table 1.
|
Changes (means, SD) in respiratory rate (RR;
min
1) and VT with NCPAP
are shown in Table 2. RR was similarly
decreased as NCPAP increased for the 2 devices.
VT was not changed with NCPAP for either device
but was greater for the variable-flow device at all CPAP levels
(P < .001).
|
VL as a function of NCPAP for each device are
shown in Fig 1, top. Although
VL increased with NCPAP with both systems,
VL was significantly greater overall with the
variable-flow device (P < .001). The corresponding
CL comparison is shown in Fig 1, bottom.
CL essentially was unchanged with constant-flow
NCPAP but was slightly increased at CPAP of 4 and 6 cm
H2O, relative to 0 cm H2O,
for the variable-flow device. At CPAP of 8 cm
H2O, this tendency was reversed and
CL was slightly (not significantly) lower with
variable flow.
|
Changes in WOBI and its component RWOB per milliliter of delivered VT are shown in Fig 2. Overall, both WOBI and RWOB were significantly lower with the variable-flow device compared with the constant-flow system (P < .001). WOBI with variable-flow NCPAP was lower by 28.8%, 29.0%, and 13.6% at CPAP of 4, 6, and 8 cm H2O, respectively. At CPAP of 0 cm H2O, RWOB was significantly greater with constant-flow NCPAP (P < .05), possibly indicating the effects of the larger mechanical impedance of the Inca nasal prongs. This difference, however, did not affect WOBI at 0 cm H2O.
|
| |
DISCUSSION |
|---|
|
|
|---|
NCPAP is increasingly being used as the primary method of ventilatory support in preterm infants with respiratory distress syndrome. This trend is based on the perceived advantages of this relatively noninvasive approach, including avoiding the risks and complications of intubation and mechanical ventilation. More important, compared with intubation and mechanical ventilation, NCPAP is believed to decrease the likelihood of pulmonary barotrauma and volutrauma in infants and consequently the development of chronic lung disease.12
The conventional method used to provide NCPAP uses constant-flow CPAP delivered via nasal prongs connected to a neonatal ventilator. In a newer system that provides variable-flow NCPAP, a flow driver delivers gas via specially designed nasal prongs.1 Studies comparing the efficacy of these 2 approaches remain scarce and are limited mostly to mechanical models.2 Furthermore, no studies have compared the possible effects of constant- versus variable-flow NCPAP on WOB and lung mechanics in infants. This perhaps is due to the difficulty of obtaining the relevant WOB measurements in neonates, particularly without altering the delivery of NCPAP.
In mechanical model studies, constant-flow NCPAP has been associated with increased WOB compared with both mask CPAP13 and variable-flow NCPAP.2 The design of the variable-flow NCPAP system offers several advantages during both inspiration and expiration. During inspiration, the design of the variable-flow nasal prongs results in high-velocity jet flows, allowing gas entrainment. This assists inspiration on demand, which, in turn, keeps the CPAP level constant.1,2 A less variable CPAP would be expected to improve and maintain VL recruitment with CPAP.
We recently reported that variable-flow NCPAP led to superior
VL compared with constant-flow
NCPAP.3 Such increased VL will
improve lung mechanics and, hence, may decrease WOB, provided that
parenchymal overdistention is avoided. Note that, on exhalation, the
variable-flow nasal prongs allows any excess gas flow from the CPAP
driver to be shunted away from the patient to ambient air, through an
expiratory outlet. This is in contrast to constant-flow NCPAP whereby
gas flow continues toward the nares during
exhalation.1,2,14 With constant-flow NCPAP, the patient's
expiratory effort then must overcome this excess flow, leading to
increased expiratory work.
Arguably, these advantages of variable-flow NCPAP should translate into superior clinical performance, including an improved success rate of extubation. Data supporting this were presented recently for ventilated preterm infants who were extubated prophylactically to CPAP. Here, investigators found that the rate of successful extubation was greater with the Infant-Flow system compared with constant-flow nasal or nasopharyngeal CPAP.15,16 However, studies investigating the long-term or prophylactic use of variable-flow NCPAP in the treatment of preterm infants have not been reported.
Critique of Experimental Methods
Measurement of lung mechanics and WOB during NCPAP is not possible
with standard methods, whereby flow and volume are measured at the
airway opening, without affecting the results. This is because
placement of an in-line flow-measuring device such as a
pneumotachograph will interfere with CPAP delivery, will cause additional imposed WOB, and is cumbersome and impractical to use in a
clinical setting. RIP provides an attractive alternative whereby
measurements of volume changes (and hence flow) are done distally at
the chest wall and therefore does not interfere with or alter how CPAP
is delivered to patients. Moreover, with the use of DC-coupled RIP,
careful measurements allow a simultaneous assessment of both
VL (DC component) and VT
(AC component). Measurement of changes in
VL
is not possible from pneumotachography. A possible disadvantage of RIP
is that it relies on an incomplete sampling of the thorax (1 abdominal
and 1 rib-cage band; see Methods) to infer volume changes. This may
limit its accuracy, particularly in the presence of significant chest
wall distortion. Although we cannot discount completely the possible
effects of this limitation of RIP, we believe that its importance is
secondary as 1) we carefully marked the location of RIP bands to ensure
identical placement, and 2) the devices were tested in random order
with each patient serving as his or her own control.
In this study, performed in preterm neonates with minimal lung disease, we showed that, compared with constant-flow NCPAP, a variable-flow device is able to recruit more VL (Fig 1, top) and increase the effective CL (Fig 1, bottom) and that it is able to do this with relatively decreased WOB (Fig 2). The latter is due partly to the larger RWOB with the constant-flow device when no CPAP flow is provided (ie, CPAP of 0 cm H2O). This indicates the presence of additional imposed WOB due to the Inca prongs (which were used exclusively in this study with the constant-flow device). Note that other commercially available nasal prongs used with conventional NCPAP may lead to less (or more) imposed WOB, depending on their specific design.
As mentioned above, lung recruitment was noticeably increased for 6 and
8 cm H2O variable-flow NCPAP (Fig 1). Although the associated WOB at these NCPAP settings remained lower than for
NCPAP = 0, they did not continue to decrease relative to
NCPAP = 4 (Fig 2). In fact, despite greater
VL, WOB was slightly
albeit not
significantly
greater than for NCPAP = 4. Given that 1) gas compression compliance is necessarily increased at higher
VL and 2) RWOB was not increased as variable-flow
NCPAP increased (Fig 2), our data might point to the possibility that
lung overdistension at the higher NCPAP levels (or equivalent
VL) may have occurred in some patients with the
variable-flow system. This is perhaps an indication of the increased
efficacy of maintaining a constant airway pressure (and hence
VL recruitment) with variable-flow NCPAP.1,2 Therefore, caution is advisable when titrating
toward the higher NCPAP levels with this device.
| |
CONCLUSION |
|---|
|
|
|---|
Variable-flow nasal CPAP potentially can provide superior respiratory support in neonates that is characterized by 1) greater and more stable volume recruitment, 2) improved CL, and 3) decreased WOB. Our findings, however, are limited to its short-term use in 24 neonates with mild lung disease or apnea. Similar investigation of the use of 1) variable-flow nasal CPAP in a larger number of neonates, 2) for longer periods of time, 3) in comparison with other nasal prong designs/constant-flow NCPAP, and 4) in infants with more significant lung disease are needed for confirmation of the reported findings.
| |
ACKNOWLEDGMENTS |
|---|
This work was supported in large part by a grant from the Cooper Faculty Practice Foundation. Hamilton Medical, Inc (Reno, Nevada) provided the variable-flow nasal CPAP device, manufactured by EME Ltd (Brighton, England).
We thank Gerald K. Arnold, PhD, for help in statistical analysis.
| |
FOOTNOTES |
|---|
Received for publication Dec 7, 2000; accepted Feb 8, 2001.
This work was presented in part at the Pediatric Academic Societies' Annual Meeting; May 1-4, 1999; San Francisco, CA.
Reprint requests to (R.H.H.) Mercy Children's Hospital, 2213 Cherry St, ACC 309, Toledo, OH 43608. E-mail: robert_habib{at}mhsnr.org
| |
ABBREVIATIONS |
|---|
NCPAP, nasal continuous positive airway pressure;
VL, lung volume;
WOB, work of breathing;
RIP, respiratory
inductance plethysmography;
Pes, esophageal pressure;
VT, tidal ventilation;
VL, change in lung volume;
CL, lung compliance;
DC, direct current;
RR, respiratory
rate;
SE, standard error.
| |
REFERENCES |
|---|
|
|
|---|
- Moa G, Nilsson K, Zetterstrom H, Jonsson LO A new device for the administration of nasal constant positive airway pressure in the newborn: an experimental study. Crit Care Med 1988; 16:1238-1242 [Medline]
- Klausner JF, Lee AY, Hutchison AA Decreased imposed work with a new nasal continuous positive airway device. Pediatr Pulmonol 1996; 22:188-194 [CrossRef][Medline]
-
Courtney SE,
Pyon KH,
Saslow JG,
Arnold GK,
Pandit PB,
Habib RH
Lung
recruitment and breathing pattern during variable versus continuous
flow nasal continuous positive airway pressure in premature infants: an
evaluation of three devices.
Pediatrics
2001;
107:304-308
[Abstract/Free Full Text] - Locke R, Greenspan JS, Shaffer TH, Rubenstein SD, Wolfson MR Effect of nasal CPAP on thoracoabdominal motion in neonates with respiratory insufficiency. Pediatr Pulmonol 1991; 11:259-264 [Medline]
- Sandberg KL, Lindstrom DP, Krueger ED, Sundell H, Cotton RB Measurement of tidal volume during high frequency ventilation by impedance plethysmography. Pediatr Res 1988; 23:253-256 [Medline]
- Warren RH, Alderson SH Face mask application for calibration of respiratory inductive plethysmography in lambs. J Dev Physiol 1988; 10:175-178 [Medline]
- Brooks LJ, DiFiore JM, Martin RJ, CHIME Study Group Assessment of tidal volume over time in preterm infants using respiratory inductance plethysmography. Pediatr Pulmonol 1997; 23:429-433 [CrossRef][Medline]
-
Pandit PB,
Pyon KH,
Courtney SE,
England SJ,
Habib RH
Lung resistance
and elastance in spontaneously breathing pre-term infants: effects of
breathing pattern and demographics.
J Appl Physiol
2000;
88:997-1005
[Abstract/Free Full Text] - Harris TR, Wood BR. Physiologic principles. In: Goldsmith JP, Karotkin EH, eds. Assisted Ventilation of the Neonate. 3rd ed. Philadelphia, PA: WB Saunders; 1996:21-68
- Cohen J. Statistical Power Analysis for the Behavioral Sciences. Hillsdale, NJ: Lawrence Erlbaum; 1988:48-52, 274-280
- Kirk R. Experimental Design: Procedures for the Behavioral Sciences. Pacific Grove, CA: Brooks/Cole; 1995:454-473
-
Van Marter LJ,
Allred EN,
Pagano M,
Do clinical markers of
barotrauma and oxygen toxicity explain interhospital variation in rates
of chronic lung disease? The Neonatology Committee for the
Developmental Epidemiology Network.
Pediatrics
2000;
105:1194-1201
[Abstract/Free Full Text] -
Goldman SL,
Brady BP,
Dumpit FE
Increased work of breathing associated
with nasal prongs.
Pediatrics
1979;
64:160-164
[Abstract/Free Full Text] - Moa G, Nilsson K Nasal continuous positive airway pressure: experiences with a new technical approach. Acta Pediatr 1993; 82:210-211 [Medline]
- Sun SC, Tein HC Randomized controlled trial of two methods of nasal CPAP (NCPAP): flow driver vs conventional NCPAP [abstract]. Pediatr Res 1999; 45:322A
-
Roukema H,
O'Brien K,
Nesbitt K,
Zaw W
A crossover trial of infant
flow (IF) continuous positive airway pressure (CPAP) versus
nasopharyngeal (NP) CPAP in the extubation of babies
1250 grams
birthweight [abstract].
Pediatr Res
1999;
45:317A
Pediatrics (ISSN 0031 4005). Copyright ©2001 by the American Academy of Pediatrics
This article has been cited by other articles:
![]() |
T Pantalitschka, J Sievers, M S Urschitz, T Herberts, C Reher, and C F Poets Randomised crossover trial of four nasal respiratory support systems for apnoea of prematurity in very low birthweight infants Arch. Dis. Child. Fetal Neonatal Ed., July 1, 2009; 94(4): F245 - F248. [Abstract] [Full Text] [PDF] |
||||
![]() |
O Karam, C Donatiello, E Van Lancker, V Chritin, R E Pfister, and P C Rimensberger Noise levels during nCPAP are flow-dependent but not device-dependent Arch. Dis. Child. Fetal Neonatal Ed., March 1, 2008; 93(2): F132 - F134. [Abstract] [Full Text] [PDF] |
||||
![]() |
A A Hutchison and S Bignall Non-invasive positive pressure ventilation in the preterm neonate: reducing endotrauma and the incidence of bronchopulmonary dysplasia Arch. Dis. Child. Fetal Neonatal Ed., January 1, 2008; 93(1): F64 - F68. [Full Text] [PDF] |
||||
![]() |
H. Boumecid, T. Rakza, A. Abazine, S. Klosowski, R. Matran, and L. Storme Influence of three nasal continuous positive airway pressure devices on breathing pattern in preterm infants Arch. Dis. Child. Fetal Neonatal Ed., July 1, 2007; 92(4): F298 - F300. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Tsuchida, D. Engelberts, M. Roth, C. McKerlie, M. Post, and B. P. Kavanagh Continuous positive airway pressure causes lung injury in a model of sepsis Am J Physiol Lung Cell Mol Physiol, October 1, 2005; 289(4): L554 - L564. [Abstract] [Full Text] [PDF] |
||||
![]() |
A G De Paoli, R Lau, P G Davis, and C J Morley Pharyngeal pressure in preterm infants receiving nasal continuous positive airway pressure Arch. Dis. Child. Fetal Neonatal Ed., January 1, 2005; 90(1): F79 - F81. [Abstract] [Full Text] [PDF] |
||||
![]() |
F Sandri, G Ancora, A Lanzoni, P Tagliabue, M Colnaghi, M L Ventura, M Rinaldi, I Mondello, P Gancia, G P Salvioli, et al. Prophylactic nasal continuous positive airways pressure in newborns of 28-31 weeks gestation: multicentre randomised controlled clinical trial Arch. Dis. Child. Fetal Neonatal Ed., September 1, 2004; 89(5): F394 - F398. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. M. Stefanescu, W. P. Murphy, B. J. Hansell, M. Fuloria, T. M. Morgan, and J. L. Aschner A Randomized, Controlled Trial Comparing Two Different Continuous Positive Airway Pressure Systems for the Successful Extubation of Extremely Low Birth Weight Infants Pediatrics, November 1, 2003; 112(5): 1031 - 1038. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. H. Habib, K. H. Pyon, S. E. Courtney, and Z. H. Aghai Spectral characteristics of airway opening and chest wall tidal flows in spontaneously breathing preterm infants J Appl Physiol, May 1, 2003; 94(5): 1933 - 1940. [Abstract] [Full Text] [PDF] |
||||
![]() |
A G De Paoli, C Morley, and P G Davis Nasal CPAP for neonates: what do we know in 2003? Arch. Dis. Child. Fetal Neonatal Ed., May 1, 2003; 88(3): F168 - F172. [Abstract] [Full Text] [PDF] |
||||
eLetters:
Read all eLetters
- CPAP and work of breathing
- Kenneth Harkavy
- Pediatrics Online, 18 Oct 2001 [Full text]
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||










