PEDIATRICS Vol. 108 No. 2 August 2001, pp. 389-394
Aerosol Therapy With Valved Holding Chambers in Young Children: Importance of the Facemask Seal

From the * Pediatric Department, R. Sieff Hospital, Safed,
Israel; and Objective. Masks are an essential
interface between valved holding chambers (VHCs), or spacers,
and a small child's face for providing aerosol therapy. Clinical
experience suggests that many young children do not cooperate with the
VHC treatment or tolerate a mask of any kind. This might impair the
mask-face seal and reduce the dose delivered to the child. The
objective of this study was to evaluate the ability of parents to
provide a good mask-face seal in infants and toddlers using 3 masks
provided with commonly used pediatric VHCs and compare this with the
seal obtained with the Hans Rudolph pediatric anesthesia mask.
Methods. A preliminary in vitro filter study was conducted
to validate the assumption that reduced ventilation as a result of
increased facemask leak reduces the drug aerosol dose delivered to the
mouth. Facemask leak then was studied in vivo for NebuChamber,
AeroChamber, BabyHaler, and Hans Rudolph masks by measuring ventilation
with an in-line pneumotachograph while the facemask was held in place by experienced parents who were asked to demonstrate how they deliver
medication to their children without any additional instruction. Thirty
children (mean age: 3.2 ± 1.4 years) performed 4 repeat studies
with each mask. The first 10 patients performed the tests once again
within 1 month. On the second occasion, the parents were coached
continuously and encouraged to hold the mask tightly against the
child's face.
Results. The AeroChamber and Hans Rudolph masks provided
the best seal as reflected in the magnitude of the ventilation measured
through them. The NebuChamber provided the poorest seal, with 45% less ventilation than the AeroChamber and Hans Rudolph masks. There was
considerable intraindividual variability for all masks (24% to 48%);
however, the variability with the NebuChamber mask was 2-fold greater
than the other masks. All ventilatory volumes during the coached
session were significantly greater than during the uncoached session.
Variability during the coached session was significantly less (except
for the BabyHaler, which remained unchanged).
Conclusions. VHCs with masks designed for use with small
children may provide a poor seal with the face, leading to reduced or
more variable dose delivery. The facemask seal is critical for
efficient aerosol delivery to infants and young children, and this
should be stressed to parents.
Department of Medicine, McMaster University Barnett
Medical Aerosol Research Laboratory, St Joseph Hospital, Hamilton,
Ontario, Canada.
![]()
ABSTRACT
Top
Abstract
Methods
Results
Discussion
References
The development of spacers and valved holding chambers
(VHCs) as metered-dose inhaler (MDI) accessories has greatly improved inhalation treatment in young children.1,2 There are 2 main advantages to these add-on devices: 1) aerodynamic filtration: the
chamber's dimensions allow the larger particles to decelerate and
deposit in the device instead of in the mouth or throat, thus reducing
impaction of particles in the upper airway; particularly with inhaled
steroids, this greatly decreases local upper respiratory tract side
effects (hoarseness, candidiasis) and systemic absorption of
medications and improves the therapeutic ratio3; and 2)
provision of a reservoir of aerosol from which the infant or child can
breathe tidally, thus ensuring aerosol delivery on inspiration without
the need to coordinate precisely the aerosol discharge and inhalation.
The addition of a facemask allows VHCs to be used with infants who
usually are nosebreathers and are too young to breathe through a
mouthpiece.4 In pediatric practice, some VHCs, designed
specifically for infants' low tidal volume and inspiratory efforts,
have improved our ability to treat adequately even premature neonates.5 This has been achieved by means of a relatively small total volume to achieve a high concentration of small particles together with very low inspiratory valve resistance and minimal dead
space.
Although the mask interface undoubtedly is an important determinant of
the aerosol dose delivered from the VHC to the infant's or toddler's
nose and mouth, its optimal characteristics have gained relatively
little attention and comparative studies of various facemasks are
lacking. The theoretical factors that determine mask efficiency include
mask-face seal, mask dead space, compliance, resistance and seal of
inspiration and expiration valves, and the dilutional effect of air
entering the mask from exhalation ports. For example, a mask with a
large dead space or a leak around the mask will decrease the dose of
aerosol obtained from the VHC, particularly in low tidal volume
breathing infants. A good seal around the nose and mouth will ensure
that inspiration is from the device rather than from the environment.
This study evaluated the ability of parents to provide a good
mask-face seal with infants and toddlers. Rather than rigorously, controlled, and supervised laboratory conditions, we encouraged the
participating parents to imitate the technique that they usually use at
home to provide aerosol medication to their children. We measured the
air leak of the 3 commercially available facemasks that are an integral
part of 3 widely available pediatric VHCs and compared it with that of
a Hans Rudolph anesthesia mask, which was considered the standard.
The study consisted of 2 parts. The first evaluated the in vitro
relationship between air leak and the dose delivered to the mouth. The
second was a controlled, randomized, crossover clinical study in
patients.
In Vitro Study
Previous in vitro and in vivo studies showed a positive
correlation between ventilation and the aerosol dose delivered to the
mouth.6-8 These studies were conducted with a tight
mask-to-face seal, however. To validate this assumed relationship also
under the conditions used in the present study, anticipating a
less-than-optimal seal, we conducted a preliminary study. This related
ventilation through the mask to the dose of albuterol delivered to the
mouth with a poor seal. To minimize the influence of the chamber on the
delivered dose, we used the NebuChamber VHC, whose chamber (unrelated
to the sealing characteristics of the mask) is known to have the most
consistent aerosol output.9 The delivered dose was defined
as the albuterol dose deposited on a filter inserted between the
NebuChamber and the mask. A new albuterol MDI (100 µg per actuation)
was used. The canister was shaken vigorously for 10 seconds, discharged
10 times, and then tested. Testing consisted of shaking the inhaler
vigorously and then firing 1 puff into the NebuChamber. This was
followed immediately by breathing from the mask for 15 seconds while
ventilation (through the mask) was measured using a screen
pneumotachograph from a portable spirometer (Flowscreen; E. Jaeger,
Wuerzburg, Germany) inserted between the mask and the filter holder.
The total additional dead space was 26 mL. Twenty-six tests were
conducted by 1 fully cooperative healthy adult. Similar respiratory
rate and tidal volumes in each test were achieved by means of a
metronome. The major independent variable was the leak between the face
and the mask. This was accomplished by having the participant hold the mask at various distances (0-10 mm) from the face and by applying various pressures on the mask at 0 distance.
We used electrostatic filters (Pari GmbH, Starnberg, Germany), which
were shown previously to collect virtually 100% of the aerosolized
medication dose.9 This technique of dose assessment was
shown previously to be accurate in determining the dose delivered to
the patient.68-10 The filters were analyzed for
albuterol content by ultraviolet spectrophotometry, and the dose
trapped on the filter was related to the corresponding ventilation by
simple regression.
Clinical Study
Patients
Thirty-nine children with asthma or wheezing were enrolled.
Inclusion criteria included age younger than 5 years, use of any VHC
with a facemask continuously for at least 1 month following appropriate
instruction (1-on-1 educational session by a respiratory nurse or
pediatric pulmonologist), and absence of symptoms on the day of the
study. Excluded were patients who did not allow application of more
than 1 mask for more than 1 test period.
![]()
METHODS
Top
Abstract
Methods
Results
Discussion
References
Masks
The following silicone masks were studied:
- NebuChamber (Astra Draco AB, Lund, Sweden): a small, thin, relatively stiff strait-edged mask, triangularly shaped, and elongated, which must be held in a single vertical position on the face.
- BabyHaler (infant size #2; Glaxo GmbH, Germany): a round mask with rolled edge (inside diameter: 5 cm).
- AeroChamber (child mask; Trudell Medical, London; Ontario, Canada): a round mask with rolled edge (inside diameter: 5.8 cm), the valve opening from the VHC is located eccentrically with respect to the mask. The mask contains an integral exhalation valve, which was sealed for the purpose of the study (because we were interested only in ventilation through the mask, sealing the exhalation valve has no effect on this variable).
- Hans Rudolph (#2, pt #669078-6995; Hans Rudolph, Inc, Kansas City, MO): a mask used mainly for anesthesia. It has a lower flap that helps to stabilize the mask against the chin. This is an anatomically contoured mask that was used as the standard against which the performance of the other masks was compared.
All masks covered both the nose and the mouth.
Assessment of Mask Leakage In Vivo
Figure 1 shows the experimental set-up. We measured the minute ventilation through the 4 masks while parents held the mask on their child's face. During the initial phase of the clinical study, parents were instructed to hold the mask and apply it to the face just the way that they do it at home during actual treatments. No additional instruction, correction of faulty technique, or other assistance was provided.
|
For the purpose of the study, the inlet portion of the mask, usually 22 mm outer diameter, was attached directly to a screen pneumotachograph of a portable electronic spirometer, calibrated on each of the study days. To quantify ventilation through the mask, we used the maximal voluntary ventilation (MVV) function. Once the mask was applied to the face, a few seconds were allowed to stabilize respiration after which the MVV test was initiated. MVV was measured for 12 seconds. The child rested for a few seconds until the results were recorded. The pneumotachograph was reset, and then the next test was done. For each mask, at least 4 such measurements were obtained. Mask administration was randomized. Approximately 15 minutes were required to test all of the masks in each patient.
Analysis Air leakage around the mask during tidal breathing was determined indirectly by comparing the ventilated volume through each mask by means of the pneumotachograph. Thus, the better the seal to the face, the greater the measured ventilation. A leak would decrease measured ventilation. However, increased MVV without any change in the facemask seal could result simply from increased respiratory rate, for example, as a result of agitation. Thus, differences in measured ventilation might not necessarily be caused by a deficient seal. To correct for such a possibility, we compared respiratory rate (which is not affected by the integrity of the seal) between the tests from each patient. Patients whose coefficient of variation (CoV) of respiratory rate was greater than 20% (2 patients) were excluded from the analysis.
Intraindividual variability in measured ventilation was expressed as CoV. Statistical analysis was done with analysis of variance with a significance level of 95% (P < .05).| |
RESULTS |
|---|
|
|
|---|
The in vitro relationship between measured ventilation and the dose delivered to the mouth (onto the filter) using variable degrees of facemask leak is shown in Fig 2. There was a positive linear relationship between the 2 variables (r = 0.55, P = .004). The greater the ventilation, the greater the delivered dose.
|
Thirty patients (17 boys and 13 girls; mean age: 3.2 ± 1.4 years; asthma duration: 1.1 ± 1.4 years) completed the clinical study. Ventilation measured for 12 seconds, expressed as liters per minute, and the CoV are shown in Table 1.
|
The AeroChamber and Hans Rudolph masks sealed most effectively as reflected in the magnitude of the leak measured through them. The NebuChamber had the poorest seal, with ventilation of 2.89 L/min, significantly less than the AeroChamber (4.24 L/min; P < .001), Hans Rudolph (4.22 L/min; P < .001), or BabyHaler masks (3.82 L/min; P < .001). There was considerable intraindividual variability for all masks; however, the variability with the NebuChamber mask was 2-fold greater than the other masks (P < .001).
Ten patients performed the tests on 2 separate occasions (before and after additional instruction). The demographics and the baseline (uncoached session) ventilation and CoV data of these 10 patients were not significantly different from the whole group of 30 patients. All ventilatory volumes of these 10 patients during the coached session were significantly (P < .05) greater than during their previous uncoached session (Fig 3). Similarly, variability during the coached session was significantly (P < .05) less, except for the BabyHaler, which remained unchanged (Fig 4).
|
|
Of 10 enrolled infants who were younger than 2 years, only 3 cooperated fully and were able to complete at least 3 attempts with each of the masks. The others would not remain still for the 12 to 15 seconds required to complete the measurements. They became agitated and pulled the mask off their face. No patient who was older than 2 years had to be excluded as a result of noncooperation.
| |
DISCUSSION |
|---|
|
|
|---|
This study demonstrated that if there is not a good seal between the mask and the child's face, then ventilation through the mask may be reduced considerably, as shown, in particular, by the less substantial NebuChamber mask. On the basis of the results of our preliminary in vitro study and because clinically relevant drug particles, in particular the fine particle fraction (<3 µm), should follow ventilation through the mask,8 it is highly likely that this also would be the case for the dose of therapeutic aerosol.
The variability of ventilation through the masks was remarkably high. Even with the same VHC mask and in the same patient, there can be considerable variability, which would be likely to translate into marked variability in the aerosol dose delivered from VHCs to the child's mouth. Indeed, relatively large variability was demonstrated previously in studies that were conducted under well-controlled laboratory conditions in which patients were instructed to seal the mask tightly on the face and all efforts were taken to ensure this fit, either by the investigator or by a nurse who was supervising the inhalation.6,7,11 That is unlike the situation in real life. Even if parents have been instructed and thus are somewhat familiar with aerosol administration using VHC and mask, it is evident from the present study that children often do not achieve an optimally tight fit between the mask and the face. Furthermore, although both ventilation and variability improved markedly when the parents were coached to apply the mask to the face properly, variability still remained between 20% (with the AeroChamber mask) and 28% (with the BabyHaler mask).
In 2 recent studies, considerable variability in the dose of aerosol delivered via VHCs was noted under real-life conditions.12,13 When a facemask is used in a similar age group, the mean CoV of the doses delivered in these studies using a filter technique was 34% for the NebuChamber metal spacer and 37% for plastic devices, values that approximate the values for CoV of ventilation observed in our study.
Despite the large intraindividual variability, the NebuChamber mask performed significantly worse than the other 3 masks and would be likely to deliver less aerosol to the mouth of the infant. This probably is because the NebuChamber mask has a sharp, flat, and relatively firm edge, whereas the others have an inwardly rounded and very flexible edge that seems to adjust better to the facial contours of the child.
The mask is an integral and vital component of the MDI/VHC delivery system in infants and toddlers, and the facemask seal is the most problematic and least robust link in the chain of events from the MDI to the child's respiratory tract. The preliminary study confirmed the validity of using total ventilation through the VHC masks to reflect the effects of mask-face seal on the aerosol dose presented to the nose and the mouth. It is a simple and practical method for evaluating facemask seal in young children in vivo.
Given the need to measure ventilation, this study was designed to simulate, as closely as possible, aerosol delivery as undertaken at home on a daily basis in infants and toddlers. During the first part of the clinical study, apart from a request to demonstrate the usual manner of applying the facemask to the child's face as parents had been instructed previously, no additional assistance was provided to parents. This was an attempt to simulate real-world conditions at home, although we recognize that in an office setting, with their technique being observed, the parents may have been anxious and so might have performed suboptimally.
Previous studies that assessed the performance of VHCs focused on the chamber portion of the device. For example, it has been demonstrated that the metal NebuChamber eliminates electrostatic charge6,7 that develops on the walls of most plastic chambers unless the latter are washed from time to time with detergent.14 Factors related to chamber size15 and valve function16,17 also are important. However, no previous attempt has been made to evaluate the facemask seal for ensuring optimum ventilation and hence aerosol drug delivery from the chamber to the lung. The facemask seal arguably is one of the most important links in the chain of events to ensure optimum aerosol therapy in this age group, and regardless of other known factors, a poor mask seal would result in an unpredictably reduced dose to the lower respiratory tract.
Of particular interest is the fit obtained in infants. Of the 10 infants who were younger than 2 years that we tried to study, reasonable cooperation was obtained in only 3. Lack of cooperation with mask treatment in very young patients was noted previously by Noble et al18, who found that many of their patients did not accept the spacer with mask, even when asleep, and had to be withdrawn from a clinical study of inhaled budesonide. It is a common complaint of parents of infants of this age that it may be very difficult to keep a mask snugly fitted to the infant's face for more than a few seconds. Persisting with a screaming infant, as parents may do, is not a good solution as it has been shown that very little aerosol medication is deposited in the lungs under these conditions.4,19 Similarly, excessive pressure on the mask might encourage crying; a happy medium must be found.
One of the potential limitations of the study relates to the validity of a 12-second measurement of ventilation. The tidal volumes of the children who participated in the study were in the range of 50 to 200 mL.11,20 Thus, assuming an average rate of 25 breaths/min11,20 (ie, 1 breath every 2-3 seconds), the resulting 4 to 6 breaths should be enough to empty even the largest of the VHCs whose masks were used in this study. Thus, the period of time during which ventilation was measured in this study should have been adequate, representing approximately the actual time recommended for inhaling the medication. We are aware that MVV testing in young noncooperative infants may not be reproducible, and we therefore excluded noncooperative infants from the study. Thus, both the leak and the variability observed in our study may be underestimated compared with real-life situations.
Another potential concern is that application of the masks in this study was slightly different from the method used during the usual treatment at home, because instead of holding several chambers, they held the same pneumotachograph handle. The reason for the approach used was twofold: 1) we wanted to standardize the tests with the only independent variable being the mask seal, and 2) we found in pilot experiments that measuring ventilation through the mask alone not only was more convenient and practical than measuring the ventilation through the mask attached to the VHC but also allowed exclusion of the potential effect of the VHC on ventilation.
The results of the present study highlight that even with cooperative children who use VHCs with masks, it should not be assumed that poor drug efficacy or an inadequate prescribed dose is the cause when asthma control is difficult to achieve. Rather, it should be assumed until proved otherwise that aerosol delivery is inefficient and suboptimal. Thus, before the medication is changed or the dosage is increased, the caregiver should be asked to demonstrate how he or she uses the VHC with mask. The present study emphasizes the importance of educating parents that a good mask fit is all-important during aerosol treatment when using MDIs and VHCs with masks. This should be stressed not only to parents who use facemasks attached to VHCs but also to those who use facemasks with nebulizers, for which similar findings have been reported.21
With close supervision and encouragement to hold the mask tightly against the child's face, ventilation significantly improved compared with conditions under which no supervision or specific instruction was provided. Parental adherence to proper inhalation technique during treatments at home and over time may, however, diminish,22 and the correct technique therefore must be reemphasized and checked frequently during follow-up visits, particularly if asthma control is suboptimal.
| |
FOOTNOTES |
|---|
Received for publication Aug 25, 2000; accepted Feb 8, 2001.
Reprint requests to (I.A.) Box 1079, Rosh-Pina, Israel 12000. E-mail: amirav{at}canaan.co.il
| |
ABBREVIATIONS |
|---|
VHC, valved holding chamber; MDI, metered-dose inhaler; MVV, maximal voluntary ventilation; CoV, coefficient of variation.
| |
REFERENCES |
|---|
|
|
|---|
-
Amirav I,
Newhouse MT
Metered-dose inhaler accessory devices in
acute asthma. Efficacy and comparison with nebulizers: a literature
review.
Arch Pediatr Adolesc Med
1997;
151:876-882
[Abstract/Free Full Text] -
O'Callaghan C,
Barry P
Spacer devices in the treatment of asthma.
BMJ
1997;
314:1061-1062
[Free Full Text] - Salzman GA, Pyszczynski DR Oropharyngeal candidiasis in patients treated with beclomethasone diproprionate delivered by metered-dose inhaler alone and with AeroChamber. J Allergy Clin Immunol 1988; 81:424-428 [CrossRef][Medline]
- Tal A, Golan H, Grauer N, Aviram M, Albin D, Quastral MR Deposition pattern of radiolabeled salbutamol inhaled from a metered-dose inhaler by means of a spacer with mask in young children with airway obstruction. J Pediatr 1996; 128:479-484 [CrossRef][Medline]
- Fok TF, Monkman S, Dolovich M, Efficiency of aerosol medication delivery from a metered dose inhaler versus jet nebulizer in infants with broncho-pulmonary dysplasia. Pediatr Pulmonol 1996; 21:301-309 [CrossRef][Medline]
- Bisgaard H A metal aerosol holding chamber devised for young children with asthma. Eur Respir J 1995; 8:856-860 [Abstract]
-
Bisgaard H,
Anhoj J,
Klug B,
Berg E
Anon-electrostatic spacer for
aerosol delivery.
Arch Dis Child
1995;
73:226-230
[Abstract/Free Full Text] -
Everard ML,
Clark AR,
Milner AD
Drug delivery from holding chambers
with attached facemask.
Arch Dis Child
1992;
67:580-585
[Abstract/Free Full Text] - Barry PW, O'Callaghan C The output of budesonide from spacer devices assessed under simulated breathing conditions. J Allergy Clin Immunol 1999; 104:1205-1210 [CrossRef][Medline]
- Berg E, Madsen J, Bisgaard H In vitro performance of three combinations of spacers and pressurized metered dose inhalers for treatment in children. Eur Respir Dis 1988; 12:472-476
-
Agertoft L,
Pedersen S
Influence of spacer device on drug delivery to
young children with asthma.
Arch Dis Child
1994;
71:217-220
[Abstract/Free Full Text] - Janssens HM, Devadason SG, Hop WC, LeSouef PN, De Jongste JC, Tiddens HA Variability of aerosol delivery via spacer devices in young asthmatic children in daily life. Eur Respir J 1999; 13:787-791 [Abstract]
- Janssens HM, Heijnen EMEW, De Jong VM, Hop WC, De Jongste JC, Tiddens HA Variability of aerosol delivery via spacer devices in wheezy infants in daily life [abstract]. Am Rev Crit Care Res Med 1999; 159:A137
- Pierart F, Wildhaber JH, Vrancken I, Devadason SG, Le Souef PN Washing plastic spacers in household detergent reduces electrostatic charge and greatly improves delivery. Eur Respir J. 1999; 13:673-678 [Abstract]
- Barry PW, O'Callaghan C The optimum size and shape of spacer devices for inhalational therapy. J Aerosol Med. 1995; 8:303-305
-
Sennhauser FH,
Sly PD
Pressure flow characteristics of the valve in
spacer devices.
Arch Dis Child
1989;
64:1305-1307
[Abstract/Free Full Text] - Fok TF, Lam K, Chan CK, Aerosol delivery to non-ventilated infants by metered dose inhaler: should a valved spacer be used? Pediatr Pulmonol 1997; 24:204-212 [CrossRef][Medline]
-
Noble V,
Ruggins NR,
Everard ML,
Milner AD
Inhaled budesonide for
chronic wheezing under 18 months of age.
Arch Dis Child
1992;
67:285-288
[Abstract/Free Full Text] - Murakami G, Igarashi T, Adachi Y, Measurement of bronchial hyperreactivity in infants and preschool children using a new method. Ann Allergy 1990; 64:383-387 [Medline]
- Iliff A, Lee VA Pulse rate, respiratory rate and body temperature of children between two months and eighteen years of age. Child Dev 1952; 23:237-245 [CrossRef][Medline]
-
Everard ML,
Clark AR,
Milner AD
Drug delivery from jet nebulisers.
Arch Dis Child
1992;
67:586-591
[Abstract/Free Full Text] - Kamps AW, van Ewijk B, Roorda RJ, Brand PL Poor inhalation technique, even after inhalation instructions, in children with asthma. Pediatr Pulmonol 2000; 29:39-42 [CrossRef][Medline]
Pediatrics (ISSN 0031 4005). Copyright ©2001 by the American Academy of Pediatrics
This article has been cited by other articles:
![]() |
T. Lindholm, R. L. Sjoberg, C. Pedroletti, A. Boman, G. L. Olsson, A. Sund, and F. Lindblad Infants' and Toddlers' Remembering and Forgetting of a Stressful Medical Procedure J. Pediatr. Psychol., March 1, 2009; 34(2): 205 - 216. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. W. Harris and G. C. Smaldone Facial and Ocular Deposition of Nebulized Budesonide: Effects of Face Mask Design Chest, February 1, 2008; 133(2): 482 - 488. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Dubik Metered Dose Inhalers: What a Difference a Mask Makes AAP Grand Rounds, June 1, 2006; 15(6): 70 - 70. [Full Text] [PDF] |
||||
![]() |
B. K. Rubin What Does It Mean When a Patient Says, "My Asthma Medication Is Not Working?" Chest, September 1, 2004; 126(3): 972 - 981. [Full Text] [PDF] |
||||
![]() |
I Amirav, I Balanov, M Gorenberg, D Groshar, and A S Luder Nebuliser hood compared to mask in wheezy infants: aerosol therapy without tears! Arch. Dis. Child., August 1, 2003; 88(8): 719 - 723. [Abstract] [Full Text] [PDF] |
||||
![]() |
I Amirav, Y Mansour, T Tiosano, S Chamny, S Chirurg, S Oren, Z Grossman, L Kahana, E Kahan, and M T Newhouse Safety of inhaled corticosteroids delivered by plastic and metal spacers Arch. Dis. Child., June 1, 2003; 88(6): 527 - 528. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. C. Randolph Aerosol Therapy with Valved Holding Chambers in Young Children, Importance of a Face Mask Seal Pediatrics, August 1, 2002; 110(2): 456 - 456. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||





, initial
study;
, coached session; *, P < .001 NebuChamber versus all other masks during original study; **,
P < .05 original versus coached session for all
masks.