PEDIATRICS Vol. 107 No. 6 June 2001, pp. 1264-1271
,
,
From the * Department of Critical Care Medicine and Surgery,
Section of Neonatology, Careggi University Hospital, University of
Florence Medical School, Florence, Italy; Objectives. The early discharge of
neonates from hospitals makes transcutaneous measurement of total
bilirubin concentration a useful tool to monitor neonatal jaundice. The
objectives of this study were to determine whether 1) transcutaneous
bilirubin (TcB) measurement, as performed using BiliCheck (BC),
correlates with total serum bilirubin (TSB) levels, measured with
standard laboratory methods and with high-pressure liquid
chromatography (HPLC-B); 2) infant race, gestational age, postnatal
age, or body weight interferes with the measurement of TcB levels in
newborn infants; 3) the variability of the TcB measurement is
comparable to the variability of TSB measurements; and 4) TcB
measurements obtained from the forehead (BCF) and sternum (BCS)
generate comparable results.
Study Design. Newborn infants who were <28 days and >30
weeks' gestational age and who underwent tests for TSB as part of
their normal care in 6 different European hospitals were studied. A
total of 210 infants were enrolled in the study, 35 at each site. Near
simultaneous (within ± 30 minutes) blood collection for TSB and
BCF and BCS measurements were performed. TSB levels were determined by
the serum bilirubin method in use at each site, and all HPLC-B
determinations were made at the same, independent laboratory.
Results. The study group consisted of 140 white,
31 Asian, 14 Hispanic, 9 African, and another 16 newborns of different
races. The correlation coefficient (r) between BCF and
HPLC-B was 0.890 (95% confidence interval = 0.858-0.915). BCF
and BCS generated similar results (r value = 0.890 for BCF and 0.881 for BCS), even if BCS slightly overestimated (mean
error = Conclusions. Because the correlation coefficient for
HPLC-B and BCF is very similar to that found for HPLC-B and laboratory
TSB, BC could be used not only as a screening device but also as a
reliable substitute of TSB determination. At higher levels of TSB, in
which phototherapy and/or exchange transfusion might be considered, BC
performed slightly better than the laboratory. The accuracy and
precision of the TcB measurement in this study was observed to
be comparable to the standard of care laboratory test.
Department of Pediatrics
and Waisman Center, University of Wisconsin School of Medicine,
Madison, Wisconsin; § Evangelisches Waldkrankenhaus Spandau, Der
Humboldt Universität Zu Berlin, Berlin, Germany;
Imperial
College School of Medicine, Hammersmith Hospital, London, England;
¶ University Hospital Zürich, Zürich, Switzerland; # Centre
d'Hemobiologie Périnatale, Hôpital Saint Antoine, Paris,
France; and ** Maternite Regionale A. Pinard, Nancy, France.
![]()
ABSTRACT
Top
Abstract
Methods
Results
Discussion
Conclusion
References
0.04 mg/dL) and BCF slightly underestimated (mean
error = 0.96 mg/dL) in comparison with HPLC-B. Analysis of
covariance demonstrated that BC accuracy was independent of
race, birth weight, gestational age, and
postnatal age of the newborn. Receiver operating characteristic
curves were evaluated for BCF and TSB, each compared with HPLC-B. With
the use of a cutoff point for HPLC-B of 13 mg/dL (222 µmol/L) and a
cutoff of 11 mg/dL on the BCF and TSB, similar sensitivity/specificity (93%/73% for BCF, 95%/76% for TSB) were observed. The use of a cutoff point for HPLC-B of 17 mg/dL (290 µmol/L) and 14 mg/dL (240 µmol/L) for BCF and TSB also produced similar sensitivity/specificity (90%/87% for the BC and 87%/83% for TSB).
Within the past 2 years, there have been at least 3 publications that addressed the care and management of neonatal
jaundice. In the first, Gartner et al1 reported the
results of a survey of American neonatologists and pediatricians in
which they found striking differences in the practice patterns of these practitioners regarding the level of monitoring and therapy of newborn
jaundice. Later, Bhutani et al2 reported on the hour-specific bilirubin nomogram that has been proposed as a more rigorous approach to predischarge screening than the day-specific values outlined by the American Academy of Pediatrics.3 Finally, Newman and Maisels4 recently discussed the
theoretical and practical issues of developing and implementing
practice guidelines specifically as they pertain to newborn jaundice.
The recent attention given to this topic has, no doubt, emerged from
the earlier reports that jaundice is the most common diagnosis that
leads to hospital readmission of newborns within the first month of
life5-7 and the anecdotal observation that there has been
a reemergence of kernicterus in this population.
The incidence of neonatal jaundice has been reported to be between 30%
and 60% of full-term newborns8 and in nearly all
premature infants. Full-term neonates who present significant hyperbilirubinemia (total serum bilirubin [TSB] >12.9 mg/dL) range from 3.5% to 12%,9 although this varies according to feeding regimens. In contrast, it was found in a large prospective study that infants who are breastfed on demand have an incidence of
hyperbilirubinemia similar to that found in formula-fed
neonates.10 The practice of early discharge (<72 hours of
age) of healthy, term newborns is growing worldwide as health care
systems look for ways to reduce costs. Because peak TSB levels
typically occur on postnatal days 3 to 5, an effective means of
screening for and monitoring the onset of hyperbilirubinemia should
enhance the safety of this growing practice.
Another complicating factor in the management of newborn jaundice has
been the widespread reports of high variability in the laboratory
measurement of bilirubin. Studies from the United
States,11,12 the United Kingdom,13 The
Netherlands and Germany,14 and New Zealand15
all indicate that the TSB values from which management decisions are
made must be considered only approximations with a wide range of
uncertainty. Furthermore, significant differences in TSB levels from
capillary blood obtained by the heel-stick method compared with blood
obtained from a vein have been reported, although this varies according
to the TSB level.16 This observation adds additional
uncertainty to the interpretation of TSB measurements. One can only
speculate as to the differences that might be seen between umbilical
artery and vein samples and even between suprahepatic and infrahepatic
venous samples.
Fortunately, the management of newborn jaundice typically requires that
therapy begin at TSB levels that are significantly below the levels at
which kernicterus is considered an immediate threat. This safety margin
still allows for therapy to be given to a relatively small number of
patients in whom hyperbilirubinemia would resolve spontaneously to
protect those who truly benefit from it.
Transcutaneous estimation of serum bilirubin is a universal practice.
The visual inspection of the skin, sclera, and mucous membranes and
observation of the cephalocaudal progression of jaundice is commonplace
as the first indication of hyperbilirubinemia. More objective methods,
such as the icterometer, have attempted to standardize the
subjective visual assessment with modest success.17 Hannemann et al18 reported the ability to measure the light reflected from the skin and mathematically convert the data to a
bilirubin value. This technique has been implemented commercially in a
2-wavelength (460 and 520 nm) device that generates a jaundice index.19 Despite reports of high correlation to TSB levels
in some populations, this device has been limited to a screening method
by the bias that race, age, and weight have on the index. Tayaba et
al20 reported good results with a new device (Chromatics
ColorMate III, New York, NY) that compares the change in yellow
coloration of the newborn skin to a baseline skin color measurement.
The drawback to this method is that every infant born, regardless of
his or her likelihood to develop subsequent hyperbilirubinemia, would
require a baseline measurement. The additional complexity and cost of
this practice is too prohibitive for widespread use.
The subject of this study is the BiliCheck ([BC], SpectRx Inc,
Norcross, GA), a new transcutaneous bilirubin (TcB) measuring device
that uses the entire spectrum of visible light (380-760 nm) reflected
from the skin. White light is transmitted into the skin of the newborn,
and the reflected light is collected for analysis. The mathematical
isolation of the light absorption of certain interfering factors
(hemoglobin, melanin, and dermal thickness) allows the absorption of
light caused by the presence of bilirubin in the capillary beds and
subcutaneous tissue to be isolated by spectral subtraction. In theory,
this will allow an unbiased measurement that is independent of the
race, age, and weight of the newborn. Early reports by Bhutani et
al21,22 suggested favorable results with the use of the
device in a diverse population.
A direct comparison between the TcB measurement and a single laboratory
method would be dependent on the accuracy of the laboratory method.
Errors in the TSB would be interpreted as errors in the TcB. Therefore,
use of the gold standard high-pressure liquid chromatography
bilirubin (HPLC-B) is necessary to serve as the true reference
value. Individual laboratory devices tend to be consistent (low
coefficient of variation) but often are not in agreement with other
instruments, because of high interlaboratory variability.
The present multicenter study is an attempt to evaluate the pooled
performance of multiple TcB devices in a diverse international population, with multiple users, and to compare the performance with
multiple laboratory methods. The objectives of our study were to
determine whether 1) TcB measurement, as performed with the use of BC,
correlates with the TSB levels, measured with standard laboratory
methods and with HPLC-B; 2) infant race, gestational age, postnatal
age, or birth weight interferes with the measurement of TcB levels in
newborn infants; 3) the variability of the TcB measurement is
comparable to the variability of TSB measurements; and 4) TcB
measurements obtained from the forehead (BCF) and sternum (BCS)
generate comparable performances.
This study was performed on newborn infants who underwent tests
for TSB as part of their normal care in 6 different European hospitals
(Queen Charlotte and Chelsea Hospital, London, England; Maternité
Regionale A. Pinard, Nancy, France; University Hospital, Zürich, Switzerland; University Hospital, Florence, Italy;
Evangelisches Waldkrankenhaus Spandau, Berlin, Germany; and
Hôpital Saint Antoine, Paris, France). A total of 210 infants
were enrolled in the study, 35 at each site. Within 30 minutes before
or after blood collection for TSB assay, TcB measurements were
performed. The blood samples were collected by heel stick or by venous
sampling as medically indicated. Standard precautions were used to
protect the sample from exposure to light to prevent photoconversion of
bilirubin in the blood. TSB levels were determined by the laboratory
method in normal use at each site (Dade Dimension [E.I. duPont de
Nemours and Company, Wilmington, DE], Ektachem DT-60 [Eastman Kodak
Company, Rochester, NY], Ginevri Microbilimeter [Rome, Italy], Ohara
Bilirubinometer, Olympus AV600 [Olympus America, Inc, Melville, NY],
Ortho Diagnostic Systems Bilimeter [Johnson & Johnson Corporation, New
Brunswick, NJ], and Pfaff Bilimeter II [Pfaff Technik & Medizin,
Neuburg/Donau, Germany]).
HPLC
A 50-µL aliquot of the serum sample was frozen at less than
TcB
For each patient, at least 1 TcB measurement was performed on
the forehead and at least 1 on the sternum/thoracic region. Repeated
measurements were taken from each anatomic site in 111 patients to
estimate instrument precision.
The device was calibrated before each measurement according to the
manufacturer's instructions to ensure the accuracy of the measurements. This requires that a reference measurement be taken from
a calibration standard (called BiliCal), which automatically adjusts
for any changes in the performance of device components over time.
To take a measurement, the probe is positioned on the infant's skin
and 5 individual scans are taken to produce 1 measurement that is
displayed in mg/dL or µmol/L. If an erroneous measurement is taken,
then an error message is displayed and the scan must be repeated.
Infants of both sexes and of any race were included in the study,
provided that they were not more than 28 days old and were at least 30 weeks of gestational age. Patients who had known skin disorders or
patients who were receiving phototherapy or exchange transfusions were
excluded from the study.
This study was approved by the Ethics Committee of each participating
hospital, and informed consent was obtained from a parent or guardian
of each patient.
Statistics
Correlation coefficients (Pearson product moment) were
calculated with the use of linear regression techniques between BC and
laboratory TSB, BC and HPLC-B, and laboratory TSB and HPLC-B, with all
sample pairs included in the analysis.
Variability and bias of laboratory TSB values were estimated for each
laboratory and among laboratories by standard parametric techniques,
with the use of the HPLC-B value as the standardized reference value
for each sample. The same calculations were performed for TcB values
with the use of the HPLC-B and the laboratory TSB as the reference
values.
A small sample size would not find a large difference in the accuracy
of the TcB significant. Conversely, a large sample size might find a
clinically insignificant difference in the accuracy of the TcB to be
statistically significant. We hypothesized that the accuracy of the TcB
measurement after pooling the data from all sites would be comparable
to the accuracy of the pooled TSB values when each is compared with the
HPLC-B values.
The null hypothesis is that the standard deviation (SD) of the BC
measurement is >1.2 times the SD (20% higher) of the laboratory TSB
value. Rejection of the null hypothesis would allow us to accept the
alternative hypothesis, which is that the error in the TcB measurement
is within 120% of the error in the laboratory TSB measurement.
H0: BC TcB SD >1.2 Times Laboratory TSB SD
To prove that TcB measurement, as performed by BC, is as
accurate, within 20%, as laboratory TSB concentration in estimating HPLC-B, the sample size was calculated with the power set at 80% and
type I error ( The sensitivity and the specificity of BC and laboratory TSB to predict
accurately HPLC-B was estimated at a range of values and plotted on
receiver operator characteristic (ROC) curves. Such curves are useful
in analyzing a test that is continuous but needs to be dichotomized.
To determine whether the patient characteristics of race, gestational
age, postnatal age, and birth weight interfered with measurements of
TcB levels, we categorized demographic values and performed an analysis
of covariance using HPLC-B as the covariate. This method tests the null
hypothesis that the mean of errors is the same for the different
categories of the categorical variables included in the model given the
adjustment of the covariate. All statistical analyses were performed
with SAS software (SAS Institute, Cary, NC).
The demographic characteristics of the group of newborn infants
studied are given in Table 1. This group
consisted of 140 white, 31 Asian, 14 Hispanic, 9 African, and
another 16 of different races. The relationship between BCF and BCS
measurements to HPLC-B and to TSB is reported in Table
2. The correlation coefficient
(r) for the combined BCF data was 0.890 (95% confidence interval [CI] = 0.858-0.915) compared with HPLC-B with an offset near 0 (0.17) and a slope near 1 (1.07). The combined BCS data also had
a high correlation to HPLC-B (R = 0.881; 95% CI = 0.846-0.908). BCS very slightly overestimated (mean error = TABLE 1 TABLE 2
![]()
METHODS
Top
Abstract
Methods
Results
Discussion
Conclusion
References
20°C and shipped on dry ice in an insulated shipping container to
the laboratory of Dr Glenn Gourley at the University of Wisconsin School of Medicine for HPLC-B determination according to the method described by Bhutani et al.22 The HPLC technician was blinded to the results of the laboratory TSB and the TcB measurements.
) at 0.05. With the use of a standard F test, a sample
size of 200 would allow rejection of the null hypothesis.
![]()
RESULTS
Top
Abstract
Methods
Results
Discussion
Conclusion
References
0.04 mg/dL) and BCF slightly underestimated (mean error = 0.961 mg/dL) in comparison with HPLC-B.
Demographic Characteristics of Newborn Infants Studied
Relationship Between Transcutaneous Bilirubin Measurements (Forehead
and Sternum) and Serum Bilirubin Concentration, Measured With HPLC
Method or With Standard Laboratory Methods (N = 210)
The correlation between BCF and laboratory TSB was very good
(r = 0.870; 95% CI = 0.832-0.899), but in this
case the offset was 1.3 with a slope near 0.89. The mean error
(TSB
BCF) was 0.13 mg/dL; this means that BCF measurement
insignificantly underestimated in comparison with laboratory TSB.
The correlations between BCS and HPLC-B and between BCS and laboratory
TSB also were high. In both cases, the mean error was negative (
0.043
and
0.865), which means that BCS, in this study, overestimated in
comparison with HPLC-B and laboratory TSB.
The intradevice coefficient of variation for the BCF was 6.69%. Because of the limitations of the serum sample size, repeated measurements for determination of intralaboratory and interlaboratory variability were not possible.
The correlation between HPLC-B and standard laboratory methods also was
high (r = 0.927; 95% CI = 0.906-0.944) even when
different methods were used in different hospitals. The mean error
(HPLC-B
TSB) was positive, and with the exception of 1 site,
TSB determined by the laboratory underestimated in comparison with
HPLC-B. Regression plots of BCF, HPLC-B, and TSB are shown in Fig
1A-C. Error plots (according to Bland
and Altman23) for each pair are shown in Fig 1D-F.
|
In Fig 2A-C, the ROC curves are reported. As the sensitivity and the specificity of a test increases, the ROC curve will appear in the upper left-hand corner of the plot. Table 3 reports the sensitivity, specificity, positive predictive value, and negative predictive value of BCF and laboratory TSB methods in relationship with HPLC-B at various clinically relevant cutoff points. To assess the clinical utility of the BC compared with the laboratory methods, we examined the ROC curves to compare the ability of the 2 methods to identify accurately the patients with TSB levels of interest. The first cutoff point selected was an HPLC-B level of 13 mg/dL (222 µmol/L). At increments of 1 mg/dL, it can be seen that a cutoff of 11 mg/dL on the BCF has approximately the same sensitivity and specificity (93%, 73%) as the laboratory TSB (95%, 76%).
|
|
At the higher levels of TSB, at which phototherapy and/or exchange transfusion might be considered, the BC performed slightly better than the laboratory. When the HPLC-B was set at 17 mg/dL (290 µmol/L), use of a cutoff point of 14 mg/dL (240 µmol/L) produced similar sensitivities and specificities: 90%, 87% for the BCF and 87%, 83% for the laboratory TSB.
The analysis of covariance (Table 4) indicates that none of the categorical variables tested are significant contributors to the BCF error when the HPLC-B is included in the model (gestational age: P = .127; birth weight: P = .155; postnatal age: P = .208; race: P = .436). Hence, the mean error of BCF compared with the mean HPLC-B level is independent of race, gestational age, postnatal age, and birth weight at measurement time. Although race was not a significant variable in the analysis of covariance, it must be pointed out that the majority of the patients (66.7%) were white and only 4.3% were of African descent. However, our finding of racial independence of the BC TcB measurement also has been demonstrated in another study.22
|
The SD of errors (compared with HPLC-B) for the BCF (SD = 2.22 mg/dL) and for the laboratory TSB (SD = 1.84 mg/dL) were tested for significance by an F test (P = .53). The nonsignificance of this test means that we cannot reject the null hypothesis; therefore, the difference between BCF variability and the TSB variability was not less than 20%. However, the clinical importance of the 0.3 mg/dL (5 µmol/L) difference between the 2 methods is subjective.
| |
DISCUSSION |
|---|
|
|
|---|
Kernicterus, which was thought to have almost completely disappeared, is now of greater concern for neonatologists and pediatricians because the earlier discharge from the hospital of mothers and neonates prevents an adequate monitoring of jaundice. The possibility of using a noninvasive, painless, and reliable method to determine the bilirubin level and its increment in the first 36 to 48 hours after birth could be very important in prevention of kernicterus. Since the early 1980s, a device dedicated to the bilirubin measurement has been proposed.19 Unfortunately, this device seems to have various limitations. First, it gives an index of jaundice, not the value of serum bilirubin concentration in standard clinical units of measurement. Race, gestational age, and body weight are factors that interfere with the accuracy of the jaundice index. However, as a screening device, this apparatus still is used in some locations. Recently, a new computer-driven, handheld device to estimate serum bilirubin from skin color of neonates was proposed. Tayaba et al20 reported good results with this new instrument, which estimates the bilirubin concentration with the use of a color discrimination algorithm. However, this instrument requires an initial set of skin color measurements within the first 30 hours after birth. Hence, to be successful with subsequent measurements, every newborn infant would require this transcutaneous measurement within the first 30 hours after birth.
BC, which measures the transcutaneous serum bilirubin by determining the intensity of specific wavelength bands that are reflected from the skin,21 acts independent of the age of the neonate. In addition, it does not seem to be influenced by the race, birth weight, gestational age, and postnatal age of the newborn for the range of patients included in this study.
The TcB determination with the use of BC seems to be more comparable to HPLC-B determination when performed on the forehead than on the sternum. In addition, the correlation between BCF and HPLC-B is slightly greater than with the laboratory TSB (0.890 vs 0.870), and it is comparable to the correlation between the laboratory TSB and the HPLC-B (0.890 vs 0.927).
Although HPLC is considered the gold standard and these results are comparable to previous reports,22 a number of factors may have contributed to degradation in both the BC and the laboratory performance in this study. Different laboratory methods were used in each hospital that participated in this study. In addition, in some hospitals, TSB was determined in the central laboratory by automated methods, whereas in others it was performed in the nursery by neonatologists or nurses. The additional handling of the serum samples, extended storage periods, and transatlantic transport of the samples could alter significantly the integrity of some samples.
In this study, both the BCF measurement and the laboratory TSB underestimated slightly the HPLC-B. Because photo or thermal degradation of a sample would normally produce a reduction in the HPLC-B determination, an alternative explanation is a concentrating effect as a result of evaporation or sublimation. However, one would not expect to see a 10% reduction in the volume of a frozen serum sample during this interval.
Conversely, BCS overestimated in comparison with HPLC-B, but the differences are very modest. In fact, the sternum measured 0.8 to 0.9 mg/dL (on average) higher than the forehead. This is counterintuitive on the basis of the observation of cephalocaudal progression of jaundice.24 A possible explanation for this observation is the effect of natural phototherapy on the forehead, which typically is exposed to ambient light more than the sternum is.
The correlation coefficient for HPLC-B and BCF (0.890) is very similar to that found for HPLC and laboratory (0.927), with a slope of the regression line of 1.07 and y intercept of 0.167. This implies that BC could be used not only as a screening device but also as a reliable substitute of TSB determination in the serum. Moreover, BC seems to have a good coefficient of variation and an acceptable accuracy.
The TSB level at which therapeutic decisions would be made depends on many factors, most notably the gestational and postnatal ages. To assess the clinical utility of the BC compared with the laboratory methods, we examined the ROC curves to compare the ability of the 2 methods to identify accurately the patients with TSB levels of interest. The first cutoff point selected was an HPLC-B level of 13 mg/dL (222 µmol/L). At increments of 1 mg/dL, it can be seen that a cutoff of 11 mg/dL on the BC has approximately the same sensitivity and specificity (93%, 73%) as the laboratory TSB (95%, 76%). Because TSB levels of <13 mg/dL account for a high percentage of the serum bilirubin analyses,25 the vast majority of invasive blood tests can be avoided depending on the population and the clinical situation (intensive care, normal newborn nursery, or outpatient facility).
At the higher levels of TSB, at which phototherapy and/or exchange transfusion might be considered, the BC performed slightly better than the laboratory. When the HPLC-B was set at 17 mg/dL (290 µmol/L), use of a cutoff point of 14 mg/dL (240 µmol/L) produced similar sensitivities and specificities: 90%, 87% for the BC and 87%, 83% for the laboratory.
| |
CONCLUSION |
|---|
|
|
|---|
TSB remains the standard of care for assessing newborn jaundice, and substitution of a new method requires substantial investigation and evidence of its superiority. The accuracy and the precision of the TcB measurement in this study was observed to be comparable to the standard of care laboratory test. Because the correlation coefficient for HPLC-B and BCF is very similar to that found for HPLC-B and laboratory TSB, BC could be used not only as a screening device but also as a reliable substitute of TSB determination.
The 2 methods do not, in fact, measure the same parameter. Although the laboratory method measures only the bilirubin that is circulating in the blood, the TcB measures the amount of bilirubin that has moved from the serum into the tissue. If this were indicative of the serum bilirubin levels that also were available to move into the brain tissue, which is our real concern, then it may offer additional information from which clinical management decisions can be made. These questions, as well as the effects of phototherapy and exchange transfusion on the TcB measurement, the effects of drugs, and the accuracy in very low birth weigh neonates and neonates <30 weeks' gestational age, all need additional study.
| |
ACKNOWLEDGMENTS |
|---|
This study was supported in part by SpectRx Inc, Norcross, Georgia.
We acknowledge the contributions made by Scott Kerr, SpectRx Inc (Norcross, GA), for his continuous support in coordinating the European trial; and by Bill Kreamer, BS, at the University of Wisconsin Waisman Center for the HPLC analysis of all serum samples and for his contribution to the technical section of the article. We also acknowledge Dr Claire Hubert of Maternité Regionale A. Pinard (Nancy, France) and Dr Daniele Piva and Dr Giovanna Bertini, Careggi University Hospital of Florence, Italy, for their great scientific and technical assistance.
| |
FOOTNOTES |
|---|
Received for publication Apr 3, 2000; accepted Sep 18, 2000.
Reprint requests to (F.F.R.) Section of Neonatology, Careggi University Hospital, Viale G. Morgagni 85, 50134 Firenze, Italy.
| |
ABBREVIATIONS |
|---|
TSB, total serum bilirubin; BC, BiliCheck; TcB, transcutaneous bilirubin; HPLC-B, high-pressure liquid chromatography bilirubin; BCF, BiliCheck forehead; BCS, BiliCheck sternum; SD, standard deviation; ROC, receiver operating characteristic; CI, 95% confidence interval.
| |
REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
D De Luca, E Zecca, M Corsello, E Tiberi, C Semeraro, and C Romagnoli Attempt to improve transcutaneous bilirubinometry: a double-blind study of Medick BiliMed versus Respironics BiliCheck Arch. Dis. Child. Fetal Neonatal Ed., March 1, 2008; 93(2): F135 - F139. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Keren, X. Luan, S. Friedman, S. Saddlemire, A. Cnaan, and V. K. Bhutani A Comparison of Alternative Risk-Assessment Strategies for Predicting Significant Neonatal Hyperbilirubinemia in Term and Near-Term Infants Pediatrics, January 1, 2008; 121(1): e170 - e179. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. B. Kolman, K. M. Mathieson, and C. Frias A Comparison of Transcutaneous and Total Serum Bilirubin in Newborn Hispanic Infants at 35 or More Weeks of Gestation J Am Board Fam Med, May 1, 2007; 20(3): 266 - 271. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Wong, D. K. Stevenson, C. E. Ahlfors, and H. J. Vreman Neonatal Jaundice: Bilirubin Physiology and Clinical Chemistry NeoReviews, February 1, 2007; 8(2): e58 - e67. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Maisels Historical Perspectives: Transcutaneous Bilirubinometry NeoReviews, May 1, 2006; 7(5): e217 - e225. [Full Text] [PDF] |
||||
![]() |
K. Grohmann, M. Roser, B. Rolinski, I. Kadow, C. Muller, A. Goerlach-Graw, M. Nauck, and H. Kuster Bilirubin Measurement for Neonates: Comparison of 9 Frequently Used Methods Pediatrics, April 1, 2006; 117(4): 1174 - 1183. [Abstract] [Full Text] [PDF] |
||||
![]() |
S Thayyil and L Marriott Can transcutaneous bilirubinometry reduce the need for serum bilirubin estimations in term and near term infants? Arch. Dis. Child., December 1, 2005; 90(12): 1311 - 1312. [Full Text] [PDF] |
||||
![]() |
G. R. Gourley, Z. Li, B. L. Kreamer, and M. R. Kosorok A Controlled, Randomized, Double-Blind Trial of Prophylaxis Against Jaundice Among Breastfed Newborns Pediatrics, August 1, 2005; 116(2): 385 - 391. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. K. Stevenson, R. J. Wong, and H. J. Vreman Reduction in Hospital Readmission Rates for Hyperbilirubinemia Is Associated with Use of Transcutaneous Bilirubin Measurements Clin. Chem., March 1, 2005; 51(3): 481 - 482. [Full Text] [PDF] |
||||
![]() |
J. R. Petersen, A. O. Okorodudu, A. A. Mohammad, A. Fernando, and K. E. Shattuck Association of Transcutaneous Bilirubin Testing in Hospital with Decreased Readmission Rate for Hyperbilirubinemia Clin. Chem., March 1, 2005; 51(3): 540 - 544. [Abstract] [Full Text] [PDF] |
||||
![]() |
Subcommittee on Hyperbilirubinemia Management of Hyperbilirubinemia in the Newborn Infant 35 or More Weeks of Gestation Pediatrics, July 1, 2004; 114(1): 297 - 316. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Ip, M. Chung, J. Kulig, R. O'Brien, R. Sege, S. Glicken, M. J. Maisels, J. Lau, and Subcommittee on Hyperbilirubinemia An Evidence-Based Review of Important Issues Concerning Neonatal Hyperbilirubinemia Pediatrics, July 1, 2004; 114(1): e130 - e153. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Maisels, E. M. Ostrea Jr, S. Touch, S. E. Clune, E. Cepeda, E. Kring, K. Gracey, C. Jackson, D. Talbot, and R. Huang Evaluation of a New Transcutaneous Bilirubinometer Pediatrics, June 1, 2004; 113(6): 1628 - 1635. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Slusher, I. A. Angyo, F. Bode-Thomas, F. Akor, S. D. Pam, A. A. Adetunji, D. W. McLaren, R. J. Wong, H. J. Vreman, and D. K. Stevenson Transcutaneous Bilirubin Measurements and Serum Total Bilirubin Levels in Indigenous African Infants Pediatrics, June 1, 2004; 113(6): 1636 - 1641. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. C. Kazmierczak, A. F. Robertson, K. P. Briley, B. Kreamer, and G. R. Gourley Transcutaneous Measurement of Bilirubin in Newborns: Comparison with an Automated Jendrassik-Grof Procedure and HPLC Clin. Chem., February 1, 2004; 50(2): 433 - 435. [Full Text] [PDF] |
||||
![]() |
M Beck, N Kau, and H Schlebusch Transcutaneous bilirubin measurement in newborn infants: evaluation of a new spectrophotometric method Arch. Dis. Child. Fetal Neonatal Ed., July 1, 2003; 88(4): F350 - F351. [Full Text] [PDF] |
||||
![]() |
V. K. Bhutani, L. H. Johnson, G. Gourley, W. D. Engle, and G. L. Jackson Measuring Bilirubin Through the Skin? Pediatrics, April 1, 2003; 111(4): 919 - 920. [Full Text] [PDF] |
||||
![]() |
C Dani, E Martelli, G Bertini, M Pezzati, L Filippi, M Rossetti, G Rizzuti, and F F Rubaltelli Plasma bilirubin level and oxidative stress in preterm infants Arch. Dis. Child. Fetal Neonatal Ed., March 1, 2003; 88(2): F119 - F123. [Abstract] [Full Text] [PDF] |
||||
![]() |
OTHER ARTICLES NOTED (Nov 01 to 18 Oct 02) Evid. Based Nurs., January 1, 2003; 6(1): e1 - 1. [Full Text] [PDF] |
||||
![]() |
T. B. Newman, P. Liljestrand, and G. J. Escobar Jaundice Noted in the First 24 Hours After Birth in a Managed Care Organization Arch Pediatr Adolesc Med, December 1, 2002; 156(12): 1244 - 1250. [Abstract] [Full Text] [PDF] |
||||
![]() |
C M Wong, P J E van Dijk, and I A Laing A comparison of transcutaneous bilirubinometers: SpectRx BiliCheck versus Minolta AirShields Arch. Dis. Child. Fetal Neonatal Ed., September 1, 2002; 87(2): F137 - 140. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. D. Engle, G. L. Jackson, D. Sendelbach, D. Manning, and W. H. Frawley Assessment of a Transcutaneous Device in the Evaluation of Neonatal Hyperbilirubinemia in a Primarily Hispanic Population Pediatrics, July 1, 2002; 110(1): 61 - 67. [Abstract] [Full Text] [PDF] |
||||
![]() |
New Device Detects Hyperbilirubinemia in Infants Journal Watch (General), June 15, 2001; 2001(615): 10 - 10. [Full Text] |
||||
| ||||||||||||||||||||||||||