PEDIATRICS Vol. 106 No. 2 August 2000, p. e17
,
,
From the * Section on Newborn Pediatrics, Pennsylvania Hospital,
University of Pennsylvania Health System, Philadelphia, Pennsylvania;
Department of Pediatrics, University of Wisconsin School of Medicine
and the Waisman Center on Mental Retardation and Human Development,
Madison, Wisconsin; and the § Department of Pediatrics, Northside
Hospital, Atlanta, Georgia.
| |
ABSTRACT |
|---|
|
|
|---|
Background. Jaundice in near-term and term newborns is a frequent diagnosis that may prompt hospital readmission in the first postnatal week. Hyperbilirubinemia, when excessive, can lead to potentially irreversible bilirubin-induced neurotoxicity. Predischarge risk assessment (at 24-72 hours of age) for subsequent excessive hyperbilirubinemia is feasible by a laboratory-based assay of total serum bilirubin (TSB).
Hypothesis. Noninvasive, transcutaneous, point-of-care measurement of transcutaneous bilirubin (TcB) predischarge by multiwavelength spectral analysis, using a portable BiliCheck device (SpectRx Inc, Norcross, GA), is clinically equivalent to measurement of TSB in a diverse, multiracial term and near-term newborn population and predictive of subsequent hyperbilirubinemia.
Methodology. We evaluated a hand-held device that uses multiwavelength spectral reflectance analysis to measure TcB (BiliCheck). The study population (490 term and near-term newborns) was racially diverse (59.1% white, 29.5% black, 3.46% Hispanic, 4.48% Asian, and 3.46% other) and was evaluated at 2 separate institutions using multiple (11) devices. The postnatal age ranged from 12 to 98 hours and the ranges of birth weights and gestational ages were 2000 to 5665 g and 35 to 42 weeks, respectively. All transcutaneous evaluations were performed contemporaneously and paired with a heelstick TSB measurement. All TSB assays were performed by high performance liquid chromatography, as well as by diazo dichlorophenyldiazonium tetrafluoroborate techniques.
Results. TSB values ranged from .2 to 18.2 mg/dL (mean ± standard deviation: 7.65 ± 3.35 mg/dL). The overall correlation of TSB (by high performance liquid chromatography technique) to TcB (by BiliCheck devices) was linear and statistically significant (r = .91; r2 = .83; TcB = .84; TSB = +.75; standard error of regression line = 1.38; P < .001; n = 490 infants; 1788 samples). Similar regression statistics were evident in subset populations categorized by race (white: r = .91 [n = 289 infants]; black: r = .91 [n = 145 infants]) as well as by gestation (term: r = .91 [n = 1625 samples]; near-term: r = .89 [n = 163 samples]). Intradevice precision was determined to be .59 mg/dL (2-3 measurements per infant with 1 device; n = 210 infants; 510 samples in a separate subset). Interdevice evaluation of 11 devices determined the precision to be .68 mg/dL (2-4 devices used for measurements per patient). In 23 of 419 of the study population infants who were in the 24- to 72-hour age range, the predischarge TSB values designated them to be at high risk for subsequent excessive hyperbilirubinemia (above the 95th percentile track on the hour-specific bilirubin nomogram). For these infants, the paired BiliCheck TcB values were all above the 75th percentile track (negative predictive value = 100%; positive predictive value = 32.86%; sensitivity = 100%; specificity = 88.1%; likelihood ratio = 8.43).
Conclusions. Our data demonstrate the accuracy and
reproducibility of the predischarge BiliCheck measurements in term and
near-term newborn infants of diverse races and ethnicities.
Infants with predischarge BiliCheck values above the 75th percentile of
hour-specific TSB values on the bilirubin nomogram may be
considered to be at high risk for subsequent excessive
hyperbilirubinemia. Further studies are needed to assess the efficacy
of this technique in preterm infants, those undergoing phototherapy,
and those with TSB values of
15 mg/dL (
256
µmol/L).
Key words:
total serum bilirubin,
high performance
liquid chromatography,
transcutaneous bilirubin,
optical density,
newborn jaundice.
Jaundice in near-term and term newborns is clinically
evident in over 60% of newborns during the first week after birth.
While hyperbilirubinemia (total serum bilirubin [TSB]: >1.0
mg/dL) occurs in nearly all infants, significant hyperbilirubinemia
(TSB: >12.9 mg/dL) and excessive hyperbilirubinemia (TSB values above
the 95th percentile for age in hours) occur in only 5% to 6% of the healthy newborn population.1-5 During the 1990s, jaundice
was the most frequent reason for readmission after early hospital
discharge.6-10 Kernicterus has been described in a small
subset of these infants.11-26 For these reasons, early
identification of newborn infants at risk for developing severe
hyperbilirubinemia and possible bilirubin-induced neurologic
dysfunction has become a public health issue. Attempts to address this
issue have included documentation of clinical and epidemiologic risk
factors,27-31 development of a practice parameter by the
American Academy of Pediatrics,32 and definition of a set
of minimal standards for early discharge and subsequent
follow-up.33-39 More recently, predischarge risk-based
assessment (24-60 hours of age) for subsequent moderate/severe
hyperbilirubinemia has been reported, based on TSB by a laboratory
assay.5
An accurate noninvasive measure of TSB would be useful, because it
would minimize blood sampling for a frequent and usually benign
clinical condition. In addition, an accurate, noninvasive test might
allow cost-effective implementation of a risk-based universal
bilirubin-screening program and follow-up to prevent the unexpected
occurrence of severe hyperbilirubinemia and perhaps even kernicterus.
Thus far, noninvasive determination of TSB has relied on the clinically
valuable but imprecise visual assessment of
jaundice1,2,40-42 or the more standardized assessment of
skin color by an icterometer.43 An objective measure of
jaundice by the Minolta Jaundice Meter device (Hill-Rom/Air Shield,
Hatboro, PA), which provides a numerical index based on spectral
reflectance, has been limited by the confounding effect of skin
pigmentation.42,44,45 Our objective was to
assess whether a transcutaneous point-of-care measurement of bilirubin
by multiwavelength spectral analysis, using the portable
BiliCheck device (SpectRx Inc, Norcross, GA), would be
clinically equivalent to the measurement of predischarge TSB in a
diverse, multiracial term and near-term newborn population and
predictive of subsequent hyperbilirubinemia.
Patient Selection
The study was conducted in the well-baby nurseries of
Pennsylvania Hospital (Philadelphia, PA) and Northside Hospital
(Atlanta, GA), with institutional review board approval at each site.
In addition, institutional review board approval was obtained at the
University of Wisconsin, Madison, the site of high-performance liquid
chromatography (HPLC) bilirubin measurement. Infants were enrolled from
March 1998 to October 1998 after obtaining informed parental consent.
Near simultaneous measurements (within 30 minutes) of transcutaneous
bilirubin (TcB) and serum bilirubin, by heelstick, were made at the
time of the routine metabolic screen before the newborn's discharge.
In some instances, measurements were performed earlier, when a TSB was
ordered at the discretion of the nursery staff for clinical signs of
jaundice. In a few infants, repeat samples were performed at ages
beyond 72 hours because of prolonged hospitalization of the mother. The
age in hours at the time of sampling was recorded and designated as
hour-specific values when corrected to the nearest hour. All infants
included in this study were discharged as healthy from the well-baby
nursery. Near-term infants were either BiliCheck Device
This is a noninvasive device consisting of a light source, a
microspectrophotometer, a fiberoptic probe, and a microprocessor control circuit with firmware for analysis and interpretation of
bilirubin measurement. The device is housed in a hand-held assembly
with a fiberoptic probe strategically placed to allow for a convenient
application to the infant's forehead. The light source is triggered
after appropriate skin application is established. Multiwavelength
spectral reflectance technique allows for the device to determine the
optical densities (ODs) attributed to bilirubin, hemoglobin, and
melanin in the subcutaneous layer in the infant skin46,47
(Fig 1). Use of in vivo multiwavelength
spectral reflectance and light transport through the complex
architecture of the neonatal skin allows for TcB
determination.48 Skin color is dependent on the presence
of: 1) oxyhemoglobin and deoxyhemoglobin (OD peaks at 420 and 585 nm),
2) melanin (which exerts an OD that behaves as if it has a zero
contribution at 837 nm and increases linearly at shorter wavelengths),
3) bilirubin (OD at 460 ± 20 nm), and 4) bilirubin
photograph-isomers, carotene, and other orange-yellow chromophores.
The dermal layer of the skin alone provides its own baseline OD, and
algorithms have been developed to determine the individual impact of
these components on the spectral reflectance. Thus, the amount of
bilirubin in the skin (TcB) can be measured by using a noninvasive
Microparts microspectrophotometer (American Laubscher Corporation,
Farmingdale, NY). Based on the light scatter that emanates from, and is
then reflected back to, the fiberoptic probe (spectral range: 380-780 nm, with a resolution of 12 nm), bilirubin-related OD values, as
reported by Jacques,48 are used to calculate the TcB
algorithms. Algorithms developed from preclinical studies were
previously validated and modified in pilot studies before this study by
the bioengineering staff at SpectRx. For the duration of the study, the
clinical investigators (V.K.B., L.H.J., C.D., and S.A.) remained
blinded to both BiliCheck TcB and HPLC TSB values.
![]()
METHODS
Top
Abstract
Methods
Results
Discussion
Conclusion
References
36 weeks of gestational age
and
2000 g birth weight, or 35 weeks of gestational age and
2500 g
birth weight. No infant had clinical manifestations of sepsis, heart or
circulatory disease, respiratory distress, or clinical evidence of
hemoglobinopathy. Data obtained after the initiation of phototherapy
were excluded. No infant had undergone an exchange transfusion.

View larger version (18K):
[in a new window]
Fig. 1.
Schematic of in vivo multiwavelength of a spectral reflectance of
neonatal jaundiced skin indicating the ODs for hemoglobin, bilirubin,
and that related to melanin. Upper panel depicts the reflectance and
the slope attributed to melanin (see text); lower panel, the OD
attributed to individual components.
Collection of TcB Data
All infants were evaluated in the nursery, usually in their bassinets, with the placement of the fiberoptic probe on their forehead. A location was chosen that was distanced from the hairline and free of any bruising, local nevus, hemangioma, or melanotic patch. The infants were placed in a supine position or cradled in a lap. A disposable clean tip, which was placed in contact with the skin, covered the probe, and light pressure was applied. Contingent on the design of the probe, application of pressure was distributed evenly. All measurements were made in the morning ambient light of the nurseries (similar use of overhead fluorescent white lights). Before each measurement, the device was calibrated to a standard reference placed in direct contact with the fiberoptic probe tip. The light source in the device was triggered for 5 spectral collections that were then averaged to provide 1 TcB measurement (blinded to the investigators for the duration of the study). During this procedure, the probe was placed in approximately the same position for each spectral collection. In a quiet infant, the entire process of data sampling took ~10 to 15 seconds. Spectral reflectance data for each infant were downloaded to a laptop computer for subsequent analysis.
To assess the reliability of a TcB measurement, the intradevice error was determined for 1 single device. For each device, 2 to 3 measurements were repeated in an infant at intervals of a few minutes. To test for reliability among multiple devices, the interdevice error in TcB measurements was determined by repeating the test with 2 to 4 separate devices for each infant at the same approximate site at intervals of 2 to 3 minutes.
Measurement of TSB
Two technicians at each of the 2 institutions did all of the blood sampling by heelstick. After warming of the heel and lancet puncture incision, blood was collected by the drip method.49 Serum was separated and divided into equal aliquots for bilirubin assay. One sample was measured locally at each hospital for clinical use. The other was placed in an amber tube, frozen immediately, and shipped in dry ice to the University of Wisconsin, Madison for multiple assays of bilirubin, including HPLC. Specific care was taken to avoid exposure to light, thus preventing photoconversion and minimizing atmospheric exposure and evaporation.
HPLC Technique to Assay Bilirubin
Serum was transferred to a .22-µm nylon microfuge centrifuge
filter (MSI, Westboro, MA), centrifuged (6370 × g for 2 minutes) and 20 µL of filtered serum was analyzed for bile pigments using reverse phase HPLC. The HPLC system consisted of a Hewlett Packard (Brookfield, WI) diode array detector, chemstation, and pump with other
details of the method as previously described.50 An MC-18
reverse phase Chromegabond, 60
pore size, 300 mm × 4 mm,
10-µm particle size, column (ES Industries, Malton, NJ), with a
Sentry µBondapak C18 125
, 10-µm 20 mm × 3.9 mm guard
column (Milipore/Waters, Milford, MA) was used. A .1-M sodium acetate (pH 4.0)/acetonitrile gradient was used at a flow rate of 2 mL/minute. This overall gradient consisted of the following linear gradients: 75%
sodium acetate for 1 minute, decreasing to 68% over 5 minutes, then
decreasing to 66% over 4 minutes, then decreasing to 65% over 3 minutes, then decreasing to 59% over 5 minutes, then decreasing to 8%
over 8 minutes and holding at 8% for 10 minutes, then changing to
methanol/dimethylsulfoxide (95:5) over 1 minute and holding there for 8 minutes, then returning to sodium acetate/acetonitrile (75:25) over 2 minutes and equilibrating at these conditions for 10 minutes before the
next analysis. Bilirubin, bilirubin conjugates, and photoproduct
detection were conducted at 436 nm (bandwidth: 4 nm) minus a reference
of 554 nm (bandwidth: 22 nm). Because the molar extinction coefficients
for bilirubin, bilirubin diglucuronide, and bilirubin monoglucuronide
are nearly identical,50,51 the standard curve derived for
bilirubin can be used to quantify all 3 bile pigments. A standard curve
was made using bilirubin standards from Sigma Chemicals (St Louis, MO;
19.9 mg/dL). The Sigma bilirubin concentration was determined by Sigma
spectrophotometrically, based on a molar extinction coefficient of
alkaline azobilirubin at 600 nm of 73 000
Lmol
1
cm
1. This lot of standard
was over 99% unconjugated bilirubin (based on total integrated area).
The initial standard curve was constructed with 6 separate injections
of bilirubin at a concentration of 19.9 mg/dL, with the resulting line
going through the origin. This standard reference curve was
subsequently verified on multiple occasions with various injections of
other Sigma bilirubin standards ranging in concentration from ~4
through 20 mg/dL. The equation for the regression line for the HPLC
bilirubin values versus the Sigma bilirubin standards was
y = .994x
.016, based on 65 different paired
measurements. Approximately 8% of the samples received were reanalyzed
to test for sample degradation during the time spent in the
autoinjector before analysis. This was done by analyzing the samples in
batches of 12 and by repeating the first sample after the 12th sample
was analyzed. No positive or negative bias was demonstrated related to
the duration of time spent in the autoinjector before analysis. This
duplicate measurement was not used for analytic purposes other than
assessment of possible sample degradation.
Data and Statistical Analysis
Care was taken that similar clinical protocol, study, reference and sample collection methods, and patient enrollment strategies were prospectively maintained, so that the data from each site could be pooled. Particular care was taken to keep the technicians, clinicians, and investigators at data collection sites blinded to the TcB data and the HPLC TSB data.
Distribution of TSB (HPLC) values was determined by plotting on the previously reported hour-specific bilirubin nomogram.5 TSB values for routine clinical use, also meticulously collected and handled by research technicians, were measured by several different commercially available assays. These assays included Ecktachem (Vitros N-Bil, Johnson and Johnson, NJ), American Optical/Leica (Buffalo, NY), and dichlorophenyldiazonium tetrafluoroborate (DPD) diazo Bilirubin-DPD reagent (Roche Diagnostics-Hitachi, Inc, Tokyo, Japan) methodologies, were all separately well-correlated to HPLC TSB values as reported elsewhere.55
The correlation of TSB (HPLC) values to TcB measurements for all enrolled infants was performed by Pearson linear regression analysis (SPSS, Chicago, IL). The data were then stratified for racial identity (as indicated by parents) and for gestational age (as determined by postnatal examination). Correlation between TSB (HPLC) and TcB for each data subset was also determined by linear regression analysis. Because linear correlation of 2 clinical measurements can be often misleading, we determined the limits of agreement that could be applied to the whole population by the statistical analysis described by Bland and Altman.52 The error distribution between the TSB (HPLC) and TcB values was evaluated for its gaussian distribution, magnitude, and confidence interval. In addition, to test the accuracy and precision of multiple BiliCheck devices (n = 11), correlation of TSB (HPLC) and TcB measurements for individual devices was also determined by linear regression analysis. To further determine the effect of confounding variables on the correlation results obtained in the study population, analysis of variance (ANOVA; using SPSS software for statistical analysis) was performed for gestational age, body weight, and race (self-declared by parents on admission).
Predischarge (24-72 Hours of Age) Risk Assessment
Paired BiliCheck and HPLC TSB data were plotted on the previously reported hour-specific bilirubin nomogram (based on the DPD diazo, TSB values, 1993-1997). The potential use of the BiliCheck measurement as a predictor of subsequent severe hyperbilirubinemia was assessed in a manner similar to that previously reported for predischarge TSB values.5 Infants were designated to be at high risk (>95th percentile track for age in hours) based on their predischarge TSB (HPLC) values as measured between 24 and 72 hours of age. Using the 75th percentile track of the TSB bilirubin nomogram as a predictive vector, paired BiliCheck and TSB (HPLC) values for all predischarge high-risk group infants were compared. Accuracy and predictive ability (negative and positive predictive values and the likelihood ratio) of the BiliCheck device (for TcB values above the 75th percentile track) to detect the predischarge high-risk group (>95th percentile) on the hour-specific bilirubin nomogram were calculated.53,54
Adverse Events
During the course of the study a daily log was maintained to document any adverse or untoward events related to the use of the fiberoptic probe and the BiliCheck device.
| |
RESULTS |
|---|
|
|
|---|
For this study, 517 infants were enrolled between 12 and 128 hours of age. Of these, 11 were excluded from analysis for loss of serum samples during transport (n = 8) and loss or corruption of spectral data for analysis (n = 3), and 16 were excluded for not meeting the strict enrollment criteria. A total of 490 infants received multiple evaluations with 2 to 4 different devices each, using a total of 11 separate devices. These provided a database of 1788 samples. In a separate subset of 510 samples obtained in 210 infants, repeat evaluations with the single device (2-3/infant) were measured.
Demographic breakdowns by birth weight, gestational age, and racial background are listed in Table 1 for the entire study population. There were no significant differences among the ethnic groups in either birth weight or gestational age (comparison by ANOVA). Age at the time of bilirubin measurement of the inpatient study population ranged from 18 to ~96 hours (median age: 39 hours). As seen in Fig 2, distribution according to HPLC TSB stratification were: for those with TSB <5 mg/dL, 19.3%; for TSB values 5.1 to 10 mg/dL, 58.0%; for those with TSB between 10.1 and 15 mg/dL, 21.7%; and for those >15 mg/dL, 1.1%. Distribution of hour-specific HPLC TSB values on the bilirubin nomogram (Fig 2) shows that 6.1% of the study population was in the high-risk zone (above the 95th percentile), 50.1% infants in the intermediate zone, and 43.8% in the low-risk zone (<40th percentile).
|
|
Accuracy and Precision of the BiliCheck Device
Clinical Accuracy The correlation between multiple, near simultaneous BiliCheck measurements of TcB and of TSB (HPLC) was linear and significant (r = .91; r2 = .83; TcB = .84x TSB + .75; standard error of regression line = 1.27; P < .001; n = 490 infants; and 1788 samples; Fig 3). Similar correlations were also evident when TcB values were compared with other TSB values obtained by DPD diazo methods and other techniques during this period. These data have been reported elsewhere.55
|
|
Intradevice and Interdevice Precision With the BiliCheck instrument, intradevice precision was determined to be .59 mg/dL (2-3 measurements per patient; n = 210 infants). These data are not included in the overall analysis. The evaluation of 11 different devices determined an interdevice precision of .68 mg/dL (2-4 measurements per patient; n = 1788 samples; n = 490 infants). The relationships for each of the 11 devices individually were demonstrated to be linear, similar, and significant.
Limits of agreement between the paired TSB (HPLC) and BiliCheck TcB values that determine the error distribution are shown in Fig 5. These values describe a normal gaussian distribution (as shown in the inset). The mean difference between paired HPLC and TcB values was .47 mg/dL, with a 95% confidence interval of
2.29 and +3.23 mg/dL.
|
Risk Assessment Distribution of the predischarge TSB values in the 24 to 72 hours of age range (n = 419 infants as selected from the study population of 490 infants) was 5.7% in the high-risk, 50.5% in the intermediate, and 43.8% in the low-risk zones of the hour-specific bilirubin nomogram. The predictive ability of the paired BiliCheck TcB values, using the 75th percentile track as a risk demarcator and predictive vector, is listed in Table 2. The negative predictive value of the BiliCheck device to identify the infants with hour-specific TSB <75th percentile values is 100%, whereas the positive predictive value is 32.86%; sensitivity, 100%; specificity, 88.1%; and the likelihood ratio, 8.43. None of the infants with TSB values above the 95th percentile for age had a BiliCheck measurement that was less than the 75th percentile track on the same TSB nomogram (except for a measurement with 1 of the 3 devices used). In addition, all infants with TSB values below the 40th percentile also had BiliCheck TcB values less than the 40th percentile.
|
| |
DISCUSSION |
|---|
|
|
|---|
Moderate to severe hyperbilirubinemia (>75th percentile for age in hours) in the newborn usually peaks between 3 and 7 days of age.1-5,56,57 This clinical condition is frequently asymptomatic and occurs after hospital discharge. Visual recognition of jaundice and its relationship to serum bilirubin is often inaccurate and unreliable as a predictor of severity of subsequent hyperbilirubinemia.1,2,5 Furthermore, infants who are healthy at discharge may be at risk for subsequent bilirubin-related neurological injury. Predischarge quantification of jaundice by a TSB assay or a noninvasive assessment of risk for excessive hyperbilirubinemia could help target clinical interventions. Preventive measures might simply be nutritional counseling, such as lactational support and possible feeding supplementation, or in high-risk infants, planning for more intrusive interventions, such as intensive phototherapy. It is in these contexts that technologies have been sought to achieve an objective, noninvasive measure of jaundice.
In this study, we report the clinical accuracy of an innovative measure of TcB in a racially diverse population, which takes into account the confounding effects of the skin thickness, blood content and flow, and maturity and pigmentation (melanin and other skin chromophores). All of these impact the transcutaneous measurement of bilirubin.46-48 Only data for term and near-term newborns (up to 4 days of age), without exposure to phototherapy and with measurements made solely on the forehead, are included in this report. We also demonstrate the clinical and predictive value of using predischarge BiliCheck values above the 75th percentile track to assess risk for subsequent excessive hyperbilirubinemia.
The clinical usefulness of TcB measure of TSB by this device was
evaluated by several means. First, the results of this diagnostic test
were compared and correlated with the laboratory-based gold standard
for TSB assay, the HPLC technique. Second, the study group was
comprised of infants with similar demographics for birth weight,
gestation, ethnicity (as usually observed in an urban well-baby
nursery), and TSB distribution as that of the normal population
previously reported for term and near-term infants. Third, the accuracy
of individual measurements of BiliCheck, as well as the intradevice and
interdevice precision, was demonstrated by standard statistical
techniques. Fourth, the Bland-Altman analysis of the data provided the
magnitude of error distribution between the BiliCheck device and the
HPLC gold standard. Finally, the clinical value of BiliCheck assay,
when found predischarge to be above (or below) the 75th percentile
track on the TSB nomogram,5 was able to identify all
infants with TSB values in the high-risk zone (>95th percentile track
for age). It is important here to note that the number of infants with
predischarge TSB values
15 mg/dL (1.1% of the study population) were
not sufficient to assess accuracy of BiliCheck measure in such infants
or to allow identification of any negative or positive bias. Studies
are being conducted to specifically enroll the extraordinary segment of
the newborn population with values
15 mg/dL or above the 95th
percentile after the 3rd postnatal day.
Errors in the laboratory measurement of TSB have been a perennial problem because of difficulties with bilirubin standards, protein matrix effects, and a variety of different methods for measurement.58 The interlaboratory variability of TSB measurement as reported in 1982 was 10% to 12%.59 In 1996, Vreman et al60 detailed the extent of interlaboratory and intralaboratory variability in nonresearch settings at that time. For 6 of the 14 laboratories (using Ektachem, Hitachi, and Paramax (Dade-Behring, Inc, Newark, DE) technologies), they reported values that were consistently >104% of target values, whereas 4 of the laboratories showed consistently lower values (<96% of target values). Intralaboratory variability, expressed as coefficient of variation, ranged as high as 17.2% for standards with bilirubin concentrations of 14.9 mg/dL. Interlaboratory variability was 16.5%. More recently, Gourley et al55 have provided preliminary data demonstrating a greater accuracy and lesser bias of the same clinical laboratory neonatal TSB methods, using the HPLC assay as the gold standard comparison. However, samples in this study were collected and handled in 2 institutions as a part of research study performed to evaluate the accuracy of the BiliCheck device. We also compared and confirmed a comparable accuracy of the TcB measurement with the HPLC gold standard, as well as the routine laboratory assays.
The American Academy of Pediatrics has established guidelines based on day-specific TSB measurements on heelstick blood samples, which accept a 10% to 20% range of inaccuracy in TSB measurements. From a clinician's perspective, guidelines must be practical, allow for variability in the measurement of bilirubin, and yet never underestimate the risk of bilirubin-induced neurologic dysfunction. Our study demonstrates that the BiliCheck device provides the clinician with a noninvasive technique to identify the high-risk infants when their predischarge TcB values are above the 75th percentile track on the hour-specific bilirubin nomogram. In addition, infants with TcB values below the 40th percentile are at low risk for subsequent severe hyperbilirubinemia, as previously reported.5 Thus, we can adjust the limits for our clinical decision-making to capture all at-risk and low-risk infants by using the 75th and the 40th percentile tracks as risk demarcators. Currently, we would recommend an immediate TSB measurement in those newborns identified by BiliCheck as being at high risk. In newborns with TcB values between the 40th and 75th percentile tracks (intermediate risk), TSB measurements may be considered discretionary, as influenced by clinical risk factors. Moreover, information about risk assessment may provide an accurate means of targeting infants or populations for treatment with heme oxygenase inhibitors and for assessing its treatment efficacy.
It was not the objective of this study to assess whether the TcB measurement of serum bilirubin would be cost-effective to substitute for the laboratory measurement of TSB in newborns. If universal predischarge screening of serum bilirubin is considered an effective program for risk assessment of subsequent hyperbilirubinemia, then TcB may be considered an accurate alternative. The cost-effectiveness of using this TcB alternative (predischarge and postdischarge) needs to be assessed in the field by a clinical trial. Cost comparisons should include not only direct costs, but also the cost of personnel, sample transport, and the emotional impact of heelstick sampling.61 In addition, cost-effectiveness and utility of the point-of-care BiliCheck need to be determined by an individual practitioner in the context of availability of results for decision-making. The substitution of TcB measure for TSB as a universal screen should await the results of a clinical trial for both efficacy and cost-effectiveness with the goals of reducing hospital readmission and risk of bilirubin-related neurological sequelae. The clinical value of TcB measurements will evolve not only with ongoing technological developments, but also with our understanding of the complex relationship between accumulation of bilirubin and levels of unbound and albumin-bound bilirubin in both intravascular and extravascular spaces.
| |
CONCLUSION |
|---|
|
|
|---|
Our data demonstrate the accuracy and reproducibility of bilirubin measurements by BiliCheck in healthy newborn infants of diverse races and ethnicities studied by this method of spectral reflectance analysis. The point-of-care BiliCheck measure, when evaluated predischarge (24-72 hours of age), also shows that when TcB values are at or above the 75th percentile of hour-specific TSB values by the bilirubin nomogram, the infants should be considered at high risk for developing excessive hyperbilirubinemia after discharge.
| |
ACKNOWLEDGMENTS |
|---|
We gratefully acknowledge the support of SpectRx personnel: Scott Kerr, Greg Newman, Felicia Thomas, and Keith Ignotz for their technical, bioengineering, and funding support.
At Pennsylvania Hospital, we appreciate the technical and administrative support of Chris Dalin, Mary Grous, Donna Spitz, Emidio Sivieri, Rosemary Dworanczyk, Jeffrey S. Gerdes, MD, and Soraya Abbasi, MD.
At the Waisman Center, Madison, we thank Brenda Egan for technical support.
| |
FOOTNOTES |
|---|
Received for publication Nov 22, 1999; accepted Mar 27, 2000.
Reprint requests to (V.K.B.) Pennsylvania Hospital, Newborn Pediatrics, 800 Spruce St, Philadelphia, PA 19107. E-mail: vibhut{at}pahosp.com
| |
ABBREVIATIONS |
|---|
TSB, total serum bilirubin; HPLC, high-performance liquid chromatography; TcB, transcutaneous bilirubin; OD, optical density; DPD, dichlorophenyldiazonium tetrafluoroborate; ANOVA, analysis of variance.
| |
REFERENCES |
|---|
|
|
|---|
early discharge of newborn and mother: a critical review of the literature.
Pediatrics
1995;
96:716-726
Monte Carlo modeling of light transport in multi-layered tissues.
Comput Methods Programs Biomed
1995;
47:131-146 [CrossRef][Medline]This article has been cited by other articles:
![]() |
E. Shaw and D. Grenier Prevention of kernicterus: New guidelines and the critical role of family physicians Can Fam Physician, April 1, 2008; 54(4): 575 - 576. [Full Text] [PDF] |
||||
![]() |
D De Luca, E Zecca, M Corsello, E Tiberi, C Semeraro, and C Romagnoli Attempt to improve transcutaneous bilirubinometry: a double-blind study of Medick BiliMed versus Respironics BiliCheck Arch. Dis. Child. Fetal Neonatal Ed., March 1, 2008; 93(2): F135 - F139. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Maisels Neonatal Jaundice Pediatr. Rev., December 1, 2006; 27(12): 443 - 454. [Full Text] [PDF] |
||||
![]() |
M. J. Maisels Historical Perspectives: Transcutaneous Bilirubinometry NeoReviews, May 1, 2006; 7(5): e217 - e225. [Full Text] [PDF] |
||||
![]() |
M. J. Maisels Use TcB as a screening tool for jaundiced newborns AAP News, July 1, 2004; 25(1): 9 - 13. [Full Text] [PDF] |
||||
![]() |
V. K. Bhutani and L. H. Johnson Urgent Clinical Need for Accurate and Precise Bilirubin Measurements in the United States to Prevent Kernicterus Clin. Chem., March 1, 2004; 50(3): 477 - 480. [Full Text] [PDF] |
||||
![]() |
L Briscoe, S Clark, and C W Yoxall Can transcutaneous bilirubinometry reduce the need for blood tests in jaundiced full term babies? Arch. Dis. Child. Fetal Neonatal Ed., May 1, 2002; 86(3): F190 - F192. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||