Published online May 5, 2008
PEDIATRICS (doi:10.1542/peds.2007-2708)
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow P3Rs: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when P3Rs are posted
Right arrow Alert me if a correction is posted
Services
Right arrow E-mail this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My File Cabinet
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via CrossRef
Google Scholar
Right arrow Articles by Garey, D. M.
Right arrow Articles by Finer, N. N.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Garey, D. M.
Right arrow Articles by Finer, N. N.
Related Collections
Right arrow Premature & Newborn

ARTICLE

Tidal Volume Threshold for Colorimetric Carbon Dioxide Detectors Available for Use in Neonates

Donna M. Garey, MD, Raymond Ward, BS, Wade Rich, BS, RRT, Gregory Heldt, MD, Tina Leone, MD and Neil N. Finer, MD

Division of Neonatology, Department of Pediatrics, University of California, San Diego, California

OBJECTIVE. Colorimetric carbon dioxide detectors are used for confirmation of endotracheal intubation. The colorimetric carbon dioxide detectors that are used for neonates are labeled for use with infants and small children >1 and <15 kg. The objective of this study was to determine the minimal tidal volume that causes a breath-to-breath color change on 2 colorimetric carbon dioxide detectors.

METHODS. Using an artificial-lung model, we determined the tidal volume threshold of 2 colorimetric carbon dioxide detectors (Pedi-Cap [Nellcor, Pleasanton, CA] or Mini StatCO2 [Mercury Medical, Clearwater, FL]) during ventilation with a T-piece resuscitator or neonatal ventilator. Digital video recordings of the colorimetric carbon dioxide detectors were made during 20 seconds of ventilation at each tidal volume. Seven clinicians who were blinded to the tidal volume reviewed the videos in random order and graded the color change to determine adequacy for clinical application.

RESULTS. The Mini StatCO2 tidal volume threshold was 0.83 mL, and the Pedi-Cap tidal volume threshold was 1.08 mL.

CONCLUSIONS. The lung model revealed that the tidal volume threshold for the tested colorimetric carbon dioxide detectors is less than the expected tidal volume of a 400-g infant and suggests that these devices are appropriate for use with any neonate to confirm intubation.

Key Words: infant • intubation • capnography • resuscitation • colorimetric carbon dioxide detector

Abbreviations: CO2—carbon dioxide • ETCO2—end-tidal carbon dioxide • NRP—Neonatal Resuscitation Program • VT—tidal volume


Accepted Dec 14, 2007.