1 From the Departments of Pediatrics and Nursing, The Johns Hopkins University School of Medicine, Baltimore, MD.
Objective. To determine the lowest dose of concentrated (U100) insulin that can be reproducibly delivered.
Methods. A telephone survey was used to determine current practices in major pediatric hospitals regarding the administration of low doses of concentrated insulin. A sensitive gravitometric technique was used to determine the error in measurement of low doses of U100 insulin by pediatric nurses and parents of diabetic children.
Results. A telephone survey revealed that doses as low as 0.5 or 1.0 U (5 to 10 µL) are routinely administered in pediatric hospitals. In our study of pediatric nurses, attempts to deliver 0.5, 1.0, and 2.0 U resulted in delivered doses of 0.975 ± 0.315, 1.638 ± 0.376, and 2.153 ± 0.435 U (mean ± standard deviation of the mean), respectively. The use of 0.3-mL syringes compared to 0.5-mL syringes did not improve accuracy or precision. Taken as a group, parents of children with diabetes were more accurate (mean delivered dose of 1.063 ± 0.276 for the 1-U dose), but the individual means ranged from 0.641 to 1.300 and coefficients of variation ranged from 5% to 33%. Only three of the seven parents could deliver 1.0 U with acceptable precision and accuracy.
Conclusions. When currently available syringes are used, insulin injections of less than 20 µL (2 U of U100) have an unacceptably large error. It is recommended that, in the inpatient setting, diluted insulin be used if the prescribed dose is less than 2.0 U.
Key Words: syringes medication errors insulin diabetes mellitus
Submitted on August 18, 1992
Accepted on January 20, 1993
This article has been cited by other articles:
![]() |
A. J. Erstad, B. L. Erstad, and D. E. Nix Accuracy and reproducibility of small-volume injections from various-sized syringes Am. J. Health Syst. Pharm., April 15, 2006; 63(8): 748 - 750. [Full Text] [PDF] |
||||
![]() |
K. Keith, D. Nicholson, and D. Rogers Accuracy and Precision of Low-Dose Insulin Administration Using Syringes, Pen Injectors, and a Pump Clinical Pediatrics, January 1, 2004; 43(1): 69 - 74. [Abstract] [PDF] |
||||
![]() |
M G Gnanalingham, P Newland, and C P Smith Accuracy and reproducibility of low dose insulin administration using pen-injectors and syringes Arch. Dis. Child., July 1, 1998; 79(1): 59 - 62. [Abstract] [Full Text] |
||||
![]() |
MEASUREMENT ERROR WITH LOW DOSES OF INSULIN Journal Watch (General), June 18, 1993; 1993(618): 6 - 6. [Full Text] |
||||