PEDIATRICS Vol. 78 No. 5 November 1986, pp. 844-849
This Article
Right arrow Full Text (PDF)
Right arrow P3Rs: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when P3Rs are posted
Right arrow Alert me if a correction is posted
Services
Right arrow E-mail this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My File Cabinet
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via CrossRef
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Solimano, A. J.
Right arrow Articles by Lockitch, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Solimano, A. J.
Right arrow Articles by Lockitch, G.

Pulse Oximetry Advantages in Infants With Bronchopulmonary Dysplasia

Alfonso J. Solimano MD1, John A. Smyth MD1, Tejinder K. Mann MD1, Susan G. Albersheim MD1, and Gillian Lockitch MD1

1 From the Departments of Pediatrics and Pathology, University of British Columbia, B.C. Children's Hospital, Vancouver, British Columbia, Canada

We studied 12 infants with a clinical and radiologic diagnosis of bronchopulmonary dysplasia who were oxygen dependent and older than 30 days. Simultaneous readings of hemoglobin oxygen saturation (Sao2) determined by two pulse oximeters (Nellcor 100, BTI Biox III) and transcutaneous (tc) Po2 (Sensor Medics, Transend) were correlated with Sao2 (Radiometer, OSM 2 Hemoximeter) and Pao2 (Corning 178) measured on blood from an indwelling arterial catheter. For each infant, the fractional inspiratory oxygen (Fio2) was adjusted to obtain three to five sets of data in the range of 70% to 95% Sao2. Fifty-three data points were generated and pooled for analysis. The slope of the regression line generated for the Nellcor 100 was .86; for the BTI Biox III, it was .91; and for the Sensor Medics Transend, it was .55, resulting in average errors of +2.5%, +1.0%, and –29%, respectively, when comparing corresponding transcutaneous and arterial values. When Sao2 was equal to or less than 95%, no infants were hyperoxic. These data confirm reports by others that tcPO2 values do not accurately represent Pao2 values in older infants with bronchopulmonary dysplasia. Pulse oximeters do not require user calibration, and their sensor is unheated so they will not cause skin burns. We conclude that pulse oximetry offers major advantages over tcPo2 measurements in the management of infants with bronchopulmonary dysplasia.

Key Words: pulse oximetry • bronchopulmonary dysplasia • oxygen monitoring • transcutaneous monitoring