Transillumination—A Technical Note

Steven M. Donn MD1, Lawrence R. Kuhns MD1, and Dietrich W. Roloff MD1

1 University of Michigan, Ann Arbor, MI 48109

Transillumination has become a standard diagnostic tool in the neonatal intensive care unit. Most centers now

See Images in the PEF File

utilize a fiberoptic light probe such as the Mini-Light or Omni-Source light (Med General, Minneapolis) for transillumination. We have used this device for several years and have noticed that over time there is a continual loss of fiberoptic bundles, which gradually diminishes the overall intensity of the light. This may not be visually apparent until a significant number of contiguous fibers have broken (Fig 1).