PEDIATRICS Vol. 66 No. 1 July 1980, pp. 31-36
This Article
Right arrow Full Text (PDF)
Right arrow An erratum has been published
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow E-mail this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My File Cabinet
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via CrossRef
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Pollitzer, M. J.
Right arrow Articles by Reynolds, E. O. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Pollitzer, M. J.
Right arrow Articles by Reynolds, E. O. R.

Continuous Monitoring of Arterial Oxygen Tension in Infants: Four Years of Experience with an Intravascular Oxygen Electrode

Melanie J. Pollitzer MB, BS1, Linda P. Soutter PhD1, and E. Osmund R. Reynolds MD, FRCP1

1 Departments of Paediatrics and Medical Physics and Bioengineering, University College Hospital and Medical School, London

Two hundred eighty-two umbilical artery catheters with oxygen electrodes on their tips were passed into the aortas of 268 newborn infants with severe respiratory illnesses between 1974-1977; 254 (90%) of the electrodes recorded Pao2 satisfactorily for a mean of 66 hours; 212 (75%) electrodes were still functioning well when removed, because they were no longer clinically needed, after a mean of 87 hours. The 27 electrodes that failed to record Pao2 on insertion, 42 that failed later, and 25 that temporarily malfunctioned were examined after removal from the infants. A structural cause was found which accounted for the failure or malfunction of 43 electrodes. Problems with the remaining 51 electrodes were thought often to have been due to clotting over the tip of the electrode. The main complication of catheterization was temporary impairment of the circulation to the legs in 69 (26%) infants. Autopsies were done on 69 (85%) of the 81 infants who died; 22 (32%) infants were found to have thrombi in an iliac artery, which spread in ten (15%) to the aorta. It was concluded that the intravascular electrode usually functioned satisfactorily and that the incidence of complications was no higher than that reported with ordinary umbilical artery catheters. Improvements in the performance of the catheter-tip electrode would be expected if structural causes for failure were eliminated and if means to reduce the incidence of clot formation around the catheter and electrode could be found. It is believed that intravascular electrodes cannot be replaced by skin electrodes For the initial management of severe respiratory illnesses.

Submitted on August 1, 1979
Accepted on October 31, 1979




This article has been cited by other articles:


Home page
Arch. Dis. Child. Fetal Neonatal Ed.Home page
C Morgan, S J Newell, D A Ducker, J Hodgkinson, D K White, C J Morley, and J M Church
Continuous neonatal blood gas monitoring using a multiparameter intra-arterial sensor
Arch. Dis. Child. Fetal Neonatal Ed., March 1, 1999; 80(2): 93F - 98.
[Abstract] [Full Text]