Safety in Incubators: A Hazard Now Corrected

Margery Franklin M.B., B.S.1, Ruth M. Heyn M.D.1, and Dietric W. Roloff M.D.1

1 Department of Pediatrics, University Hospital, University of Michigan, Medical Center, Ann Arbor, Michigan 48104

A recent experience made us aware of a potential hazard from equipment used in the care of newborn infants. A newborn boy who weighed 1,585 gm after 36 weeks' gestation and thus was small for date, was cared for in an Air-Shields C-86 Isolette, Intensive Care Model. He became very active and was able to change position inside the incubator. On several occasions he was seen with his head or feet pressed against the sides of the incubator.