From the Department of Pediatrics, University Hospital, Linköping and Central Hospital, Uddevalla, Sweden
--> Objective. To improve metabolic control and prevent complications, both acute and late, we need to adjust treatment on the basis of the blood glucose (BG) profile, as not even the most active BG self-monitoring gives sufficient information.
Design. We have used Continuous Glucose Monitoring System (CGMS; Medtronic MiniMed, Northridge, CA) in a controlled crossover study including 27 diabetic patients aged 12.5 ± 3.3 (mean; standard deviation; range: 519) years. All patients were treated with intensive insulin therapy, 14 with multiple injections, and 13 with pumps. The patients were randomized into an open or blind study arm. Both arms wore the CGMS sensor for 3 days every 2 weeks. CGMS profiles were used in the open study arm to adjust insulin therapy at follow-up visits every 6 weeks. Both the patients and the diabetes team were masked to the CGMS profiles in the blinded arm, and insulin therapy adjustments were based solely on 7-point BG profiles performed by the patients. At 3 months the 2 study arms were crossed over.
Results. Despite initial problems with a device new to both patients and the diabetes team, hemoglobin A1C decreased significantly in the open arm (from 7.70%7.31%) but not in the blind arm (7.75%7.65%). A total of 26/27 patients experienced daytime low subcutaneous glucose (<3.0 mmol/L; .8 episodes/day; duration 58 ± 29 minutes; 5.5% of total time), and 27/27 patients had at least 1 nocturnal episode of low subcutaneous glucose (.4 episodes/night; duration 132 ± 81 minutes; 10.1% of total time).
Conclusions. Use of CGMS facilitated an improved treatment, and patients received new insight and increased motivation. In this study, we found CGMS to be a useful tool for education and improving metabolic control.
Key Words: type 1 diabetes children glucose sensing Continuous Glucose Monitoring System metabolic control hypoglycemia Dawn phenomenon Somogyi phenomenon rebound phenomenon
Abbreviations: CGMS, Continuous Glucose Monitoring System BG, blood glucose MDI, multiple daily injection HbA1C, hemoglobin A1C DCCT, Diabetes Control and Complications Trial
This article has been cited by other articles:
![]() |
S. K. Garg, W. C. Kelly, M. K. Voelmle, P. J. Ritchie, P. A. Gottlieb, K. K. McFann, and S. L. Ellis Continuous Home Monitoring of Glucose: Improved glycemic control with real-life use of continuous glucose sensors in adult subjects with type 1 diabetes Diabetes Care, December 1, 2007; 30(12): 3023 - 3025. [Full Text] [PDF] |
||||
![]() |
P. Augstein, L. Vogt, K.-D. Kohnert, E.-J. Freyse, P. Heinke, and E. Salzsieder Outpatient Assessment of Karlsburg Diabetes Management System-Based Decision Support Diabetes Care, July 1, 2007; 30(7): 1704 - 1708. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Garg and L. Jovanovic Relationship of Fasting and Hourly Blood Glucose Levels to HbA1c Values: Safety, accuracy, and improvements in glucose profiles obtained using a 7-day continuous glucose sensor Diabetes Care, December 1, 2006; 29(12): 2644 - 2649. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Deiss, J. Bolinder, J.-P. Riveline, T. Battelino, E. Bosi, N. Tubiana-Rufi, D. Kerr, and M. Phillip Improved Glycemic Control in Poorly Controlled Patients with Type 1 Diabetes Using Real-Time Continuous Glucose Monitoring Diabetes Care, December 1, 2006; 29(12): 2730 - 2732. [Full Text] [PDF] |
||||
![]() |
K. Yates, A. Hasnat Milton, K. Dear, and G. Ambler Continuous Glucose Monitoring-Guided Insulin Adjustment in Children and Adolescents on Near-Physiological Insulin Regimens: A randomized controlled trial Diabetes Care, July 1, 2006; 29(7): 1512 - 1517. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. K. Fisher and M. Halvorson Future Developments in Insulin Pump Therapy: Progression From Continuous Subcutaneous Insulin Infusion to a Sensor-Pump System The Diabetes Educator, January 1, 2006; 32(1): 47S - 52S. [Full Text] [PDF] |
||||
![]() |
B. W. Bode, S. Schwartz, H. A. Stubbs, and J. E. Block Glycemic Characteristics in Continuously Monitored Patients With Type 1 and Type 2 Diabetes: Normative values Diabetes Care, October 1, 2005; 28(10): 2361 - 2366. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Valentine Continuous Glucose Monitoring Has Left the Station: Are You Onboard? The Diabetes Educator, September 1, 2005; 31(5): 649 - 662. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. C. Klonoff Continuous Glucose Monitoring: Roadmap for 21st century diabetes therapy Diabetes Care, May 1, 2005; 28(5): 1231 - 1239. [Full Text] [PDF] |
||||
![]() |
D. E. Goldstein, R. R. Little, R. A. Lorenz, J. I. Malone, D. Nathan, C. M. Peterson, and D. B. Sacks Tests of Glycemia in Diabetes Diabetes Care, July 1, 2004; 27(7): 1761 - 1773. [Full Text] [PDF] |
||||
![]() |
A. E. DeClue, L. A. Cohn, M. E. Kerl, and C. E. Wiedmeyer Use of Continuous Blood Glucose Monitoring for Animals With Diabetes Mellitus J. Am. Anim. Hosp. Assoc., May 1, 2004; 40(3): 171 - 173. [Full Text] [PDF] |
||||