This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow P3Rs: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when P3Rs are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow E-mail this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My File Cabinet
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via CrossRef
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Mahajan, L.
Right arrow Articles by Kay, M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Mahajan, L.
Right arrow Articles by Kay, M.
Related Collections
Right arrow Gastrointestinal Tract

PEDIATRICS Vol. 101 No. 2 February 1998, pp. 260-263

Reproducibility of 24-Hour Intraesophageal pH Monitoring in Pediatric Patients

Received Mar 17, 1997; accepted Jun 5, 1997.

Lori Mahajan, Robert Wyllie, Luis Oliva, Francisco Balsells, Rita Steffen, and Marsha Kay

From the Section of Pediatric Gastroenterology and Nutrition, Cleveland Clinic Foundation, Cleveland, Ohio.

Objective.  Despite the widespread use of 24-hour intraesophageal pH monitoring for evaluation of gastroesophageal reflux in infants and children, there is little published information regarding the reproducibility of ambulatory pH studies in this patient population. The purpose of our study was to evaluate the reproducibility of 24-hour intraesophageal pH monitoring in pediatric patients.

Methods.  We prospectively investigated 26 patients with symptoms suggestive of gastroesophageal reflux (14 females and 12 males) ranging in age from 1 month to 18 years (mean, 9.2 years). The patients underwent extended intraesophageal pH monitoring over two consecutive 24-hour periods.

Results.  Data analysis revealed that the overall reproducibility of ambulatory 24-hour pH monitoring is only 69% (r = 0.32). Eight of 26 patients had conflicting results on day 1 compared with results on day 2. Of the 8 patients with conflicting results on day 1 versus day 2, 5 had normal studies on day 1, but demonstrated pathologic reflux on day 2. Thus, the false-negative rate for day 1 was 19.2%. The kappa  statistic calculated for the total time the pH was abnormal was 0.32, with values <0.4 representing poor correlation. Spearman correlation coefficients indicated that the percentage of time with pH <4 (r = 0.64) and the number of reflux episodes (r = 0.71) per 24-hour period are the most reproducible pH parameters.

Conclusion.  The reproducibility of 24-hour intraesophageal pH monitoring in the pediatric population is suboptimal. The investigation should be extended or repeated if the result does not correlate with the patient's clinical history.

Key words: pH, reproducibility, children, reflux.




This article has been cited by other articles:


Home page
Arch Otolaryngol Head Neck SurgHome page
K. W. Rosbe, M. A. Kenna, and A. D. Auerbach
Extraesophageal Reflux in Pediatric Patients With Upper Respiratory Symptoms
Arch Otolaryngol Head Neck Surg, November 1, 2003; 129(11): 1213 - 1220.
[Abstract] [Full Text] [PDF]


Home page
PediatricsHome page
T. G. Wenzl, S. Schneider, F. Scheele, J. Silny, G. Heimann, and H. Skopnik
Effects of Thickened Feeding on Gastroesophageal Reflux in Infants: A Placebo-Controlled Crossover Study Using Intraluminal Impedance
Pediatrics, April 1, 2003; 111(4): e355 - 359.
[Abstract] [Full Text] [PDF]


Home page
PediatricsHome page
S. G. Jolley;, A. Sawczenko, G. Gray, B. K. Sandhu;, M. Kay, L. Mahajan, and R. Wyllie
Reproducibility of 24-Hour Intraesophageal pH Monitoring
Pediatrics, June 1, 2000; 105(6): 1371 - 1372.
[Full Text]